Propaq MD Product Functions
9650-0806-01 Rev. D
Propaq MD Operator’s Guide
1-11
Defibrillator Output Energy
Propaq MD defibrillators can deliver biphasic energy from 1 joule to 200 joules. The energy
delivered through the chest wall, however, is determined by the patient’s transthoracic
impedance. An adequate amount of electrolyte gel must be applied to the paddles and a force of
10 to 12 kilograms (22 to 26.4 pounds) must be applied to each paddle in order to minimize this
impedance. If hands-free therapy electrodes are used, make sure that they are properly applied.
(Refer to the instructions on the electrode package).
External Pacemaker
Propaq MD defibrillators include a transcutaneous pacemaker consisting of a pulse generator
and ECG-sensing circuitry. Noninvasive transcutaneous pacing (NTP) is an established and
proven technique. This therapy is easily and rapidly applied in both emergency and
nonemergency situations when temporary cardiac stimulation is indicated.
The output current of the pacemaker is continuously variable from 10 to 140 mA (the output
current is 0 mA when paused). The rate is continuously variable from 30 to 180 pulses per
minute (ppm), by increments of 5 ppm (10 bpm when greater than 100 ppm).
The pacing output pulse is delivered to the heart via ZOLL hands-free defibrillation/pacing
electrodes placed on the patient’s back and the precordium.
Proper operation of the device, together with correct electrode placement, is critical to
obtaining optimal results. Every operator must be thoroughly familiar with these operating
instructions.
ECG Monitoring
The patient’s ECG is monitored by connecting the patient to the unit via a 3-, 5-, or 12-lead
patient cable or hands-free therapy electrodes. The ECG waveform is presented on the display
along with the following information:
•
averaged heart rate, derived by measuring R to R intervals
•
lead selection - I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6 (with ECG cable),
PADDLES, or PADS.
•
ECG size - 0.125, 0.25, 0.50, 1.0, 2.0, 4.0 cm/mV, AUTO
•
status messages
The ECG bandwidth is user selectable.
Electrodes
The Propaq MD units will defibrillate, cardiovert, and monitor ECG using hands-free therapy
electrodes. The Propaq MD unit will pace using ZOLL hands-free therapy electrodes.
Energy Select
,
Charge
and
Shock
controls are located on the paddles and front panel. When
using hands-free therapy electrodes, you must use the controls on the front panel of the unit. To
switch between paddles and hands-free therapy electrodes, remove the multifunction cable
(MFC or OneStep) from the apex paddle and connect the hands-free therapy electrodes to the
cable.
Summary of Contents for Propaq MD
Page 1: ...Propaq MD Operator s Guide 9650 0806 01 Rev D...
Page 40: ...CHAPTER 1 GENERAL INFORMATION 1 28 www zoll com 9650 0806 01 Rev D...
Page 60: ...CHAPTER 2 PRODUCT OVERVIEW 2 20 www zoll com 9650 0806 01 Rev D...
Page 74: ...CHAPTER 4 TRENDS 4 4 www zoll com 9650 0806 01 Rev D...
Page 84: ...CHAPTER 5 ALARMS 5 10 www zoll com 9650 0806 01 Rev D...
Page 94: ...CHAPTER 6 MONITORING ECG 6 10 www zoll com 9650 0806 01 Rev D...
Page 104: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 0806 01 Rev D...
Page 120: ...CHAPTER 8 MONITORING NON INVASIVE BLOOD PRESSURE NIBP 8 16 www zoll com 9650 0806 01 Rev D...
Page 154: ...CHAPTER 11 MONITORING INVASIVE PRESSURES IBP 11 10 www zoll com 9650 0806 01 Rev D...
Page 160: ...CHAPTER 12 MONITORING TEMPERATURE 12 6 www zoll com 9650 0806 01 Rev D...
Page 186: ...CHAPTER 14 12 LEAD ECG INTERPRETIVE ANALYSIS 14 16 www zoll com 9650 0806 01 Rev D...
Page 212: ...CHAPTER 17 ADVISORY CPR PROTOCOL DEFIBRILLATION 17 6 www zoll com 9650 0806 01 Rev D...
Page 222: ...CHAPTER 18 EXTERNAL PACING 18 10 www zoll com 9650 0806 01 Rev D...
Page 240: ...CHAPTER 21 PATIENT DATA 21 6 www zoll com 9650 0806 01 Rev D...
Page 268: ...CHAPTER 22 COMMUNICATIONS 22 28 www zoll com 9650 0806 01 Rev D...
Page 284: ...CHAPTER 24 MAINTENANCE 24 12 www zoll com 9650 0806 01 Rev D...
Page 326: ...ACCESSORIES B 6 www zoll com 9650 0806 01 Rev D...