9650-0806-01 Rev. D
Propaq MD Operator’s Guide
24-11
Signature
_______________________________________________
Propaq MD OPERATOR’S SHIFT CHECKLIST
Date: ________________ Shift: _______________ Location: _______________
Mfr/Model No.: ____________________________ Serial No. or Facility ID No.: _________________
At the beginning of each shift, inspect the unit. Indicate whether all requirements have been met.
Note any corrective actions taken. Sign the form.
Okay as found
Corrective Action/Remarks
1. Defibrillator Unit
Clean, no spills, and casing undamaged
2. Cables/Connectors
a. Inspect for damaged insulation, frayed/broken wires, or bent
connector pins
b. Connectors engage securely
3. Sensors (pulse oximetry, NIBP cuff and hose, temperature sensors)
a. Inspect for signs of damage or excessive wear
b. Connectors engage securely
4. Paddles
a. Clean, not pitted, or damaged.
b. Switches operate freely
c. Cables and connectors free of damage and engage securely
5. Supplies and Accessories
a. Therapy pads in sealed pouches (2 sets, not expired)
b. Defib gel or gel pads
c. ECG monitoring electrodes
d. Alcohol wipes
e. Razors/scissors
f. Recorder paper
6. Batteries
a. Fully charged battery installed in unit
b. Fully charged spare battery available
7.
Ready For Use Indicator shows Ready - No flashing or Do Not Use symbol
8. Operational Checks
A. Power Up Sequence
a. Battery charge and AC indicators illuminate with AC power
b. Audio beeps heard and Visual Alarm indicators briefly illuminate
c. Self Test passed
B. Hands Free Defibrillation (Test with only battery power)
a. CHECK THERAPY ELECTRODES and PADS SHORT
DETECTED messages display
b. Charge time < 7 seconds
c. DEFIB SHORT TEST passed at 30 Joules
C. Pacer Test (Test with only battery power)
a. Printer prints Pace markers every 25 mm at 60 ppm
b. PACING: SHORT DETECTED message displays-- no error
messages at 100 mA
c. PACING: CHECK THERAPY ELECTRODES message displays
D. Paddles
a. APPLY PADDLES TO PATIENT message displays when paddles
are connected to MFC
b. Paddle switches functional (Recorder, Energy Select, Charge, Shock)
c. DEFIB SHORT TEST passed at 30 Joules
E. Reconnect Unit to AC Power
____________ Major problem(s) identified (OUT OF SERVICE)
Summary of Contents for Propaq MD
Page 1: ...Propaq MD Operator s Guide 9650 0806 01 Rev D...
Page 40: ...CHAPTER 1 GENERAL INFORMATION 1 28 www zoll com 9650 0806 01 Rev D...
Page 60: ...CHAPTER 2 PRODUCT OVERVIEW 2 20 www zoll com 9650 0806 01 Rev D...
Page 74: ...CHAPTER 4 TRENDS 4 4 www zoll com 9650 0806 01 Rev D...
Page 84: ...CHAPTER 5 ALARMS 5 10 www zoll com 9650 0806 01 Rev D...
Page 94: ...CHAPTER 6 MONITORING ECG 6 10 www zoll com 9650 0806 01 Rev D...
Page 104: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 0806 01 Rev D...
Page 120: ...CHAPTER 8 MONITORING NON INVASIVE BLOOD PRESSURE NIBP 8 16 www zoll com 9650 0806 01 Rev D...
Page 154: ...CHAPTER 11 MONITORING INVASIVE PRESSURES IBP 11 10 www zoll com 9650 0806 01 Rev D...
Page 160: ...CHAPTER 12 MONITORING TEMPERATURE 12 6 www zoll com 9650 0806 01 Rev D...
Page 186: ...CHAPTER 14 12 LEAD ECG INTERPRETIVE ANALYSIS 14 16 www zoll com 9650 0806 01 Rev D...
Page 212: ...CHAPTER 17 ADVISORY CPR PROTOCOL DEFIBRILLATION 17 6 www zoll com 9650 0806 01 Rev D...
Page 222: ...CHAPTER 18 EXTERNAL PACING 18 10 www zoll com 9650 0806 01 Rev D...
Page 240: ...CHAPTER 21 PATIENT DATA 21 6 www zoll com 9650 0806 01 Rev D...
Page 268: ...CHAPTER 22 COMMUNICATIONS 22 28 www zoll com 9650 0806 01 Rev D...
Page 284: ...CHAPTER 24 MAINTENANCE 24 12 www zoll com 9650 0806 01 Rev D...
Page 326: ...ACCESSORIES B 6 www zoll com 9650 0806 01 Rev D...