Specifications
A-2
www.zoll.com
9650-0806-01 Rev. D
Defibrillator
Charge Time
:
•
Less than 7 seconds with a new, fully charged battery (first 15 charges to 200 joules).
•
For the sixteenth discharge at maximum energy, the charge time is less than 10 seconds.
Depleted batteries result in a longer defibrillator charge time.
•
Less than 15 seconds when operating without a battery, using AC power alone at 90% of the
rated mains voltage.
•
Less than 25 seconds from the initial power on, with a new, fully charged battery pack
(depleted by up to fifteen 200 joule discharges) or when operating without a battery, using
AC power alone at 90% of the rated mains voltage.
Rhythm Analysis and Charge Time in AED Mode
•
Less than 30 seconds with a new, fully charged battery (first 15 charges to 200 joules).
•
For the sixteenth discharge at maximum energy, the analysis and charge time is less than 30
seconds. Depleted batteries result in a longer defibrillator charge time.
•
Less than 30 seconds when operating without a battery, using AC power alone at 90% of the
rated mains voltage.
•
Less than 40 seconds from the initial power on, with a new, fully charged battery pack
(depleted by up to fifteen 200 joule discharges) or when operating without a battery, using
AC power alone at 90% of the rated mains voltage.
Patient Impedance Range:
10- 300 ohms
Synchronized Mode
: Synchronizes defibrillator discharge to the patient’s R wave. SYNC is
indicated on the display with R wave markers above the ECG waveform on the screen and
stripchart. When ECG is monitored by the device, meets the DF-80:2003 requirement of 60ms
maximum time delay between the peak of the R wave and the delivery of energy.
Table A-1 shows the characteristics of the Propaq MD Rectilinear Biphasic™ waveform when
discharged into 25 ohm, 50 ohm, 100 ohm, 125 ohm, 150 ohm and 175 ohm loads at the
maximum energy setting of 200 joules.
Table A-1.
Propaq MD Rectilinear Biphasic Waveform Characteristics
200 J discharged into
25
50
100
125
150
175
First phase
Maximum initial current
31.4 A
30.4 A
19.7 A
19.4 A
16.7 A
15.6 A
Average current
27.1 A
24.9 A
17.5 A
16.2 A
14.4 A
13.2 A
Duration
6 ms
6 ms
6 ms
6 ms
6 ms
6 ms
Interphase duration
(between first and
second phases)
200
s
200
s
200
s
200
s
200
s
200
s
Second phase
Initial current
29.2 A
18.8 A
15.1 A
13.2 A
12.1 A
11 A
Average current
14.7 A
13 A
12.5 A
11.3 A
10.7 A
9.9 A
Duration
4 ms
4 ms
4 ms
4 ms
4 ms
4 ms
Summary of Contents for Propaq MD
Page 1: ...Propaq MD Operator s Guide 9650 0806 01 Rev D...
Page 40: ...CHAPTER 1 GENERAL INFORMATION 1 28 www zoll com 9650 0806 01 Rev D...
Page 60: ...CHAPTER 2 PRODUCT OVERVIEW 2 20 www zoll com 9650 0806 01 Rev D...
Page 74: ...CHAPTER 4 TRENDS 4 4 www zoll com 9650 0806 01 Rev D...
Page 84: ...CHAPTER 5 ALARMS 5 10 www zoll com 9650 0806 01 Rev D...
Page 94: ...CHAPTER 6 MONITORING ECG 6 10 www zoll com 9650 0806 01 Rev D...
Page 104: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 0806 01 Rev D...
Page 120: ...CHAPTER 8 MONITORING NON INVASIVE BLOOD PRESSURE NIBP 8 16 www zoll com 9650 0806 01 Rev D...
Page 154: ...CHAPTER 11 MONITORING INVASIVE PRESSURES IBP 11 10 www zoll com 9650 0806 01 Rev D...
Page 160: ...CHAPTER 12 MONITORING TEMPERATURE 12 6 www zoll com 9650 0806 01 Rev D...
Page 186: ...CHAPTER 14 12 LEAD ECG INTERPRETIVE ANALYSIS 14 16 www zoll com 9650 0806 01 Rev D...
Page 212: ...CHAPTER 17 ADVISORY CPR PROTOCOL DEFIBRILLATION 17 6 www zoll com 9650 0806 01 Rev D...
Page 222: ...CHAPTER 18 EXTERNAL PACING 18 10 www zoll com 9650 0806 01 Rev D...
Page 240: ...CHAPTER 21 PATIENT DATA 21 6 www zoll com 9650 0806 01 Rev D...
Page 268: ...CHAPTER 22 COMMUNICATIONS 22 28 www zoll com 9650 0806 01 Rev D...
Page 284: ...CHAPTER 24 MAINTENANCE 24 12 www zoll com 9650 0806 01 Rev D...
Page 326: ...ACCESSORIES B 6 www zoll com 9650 0806 01 Rev D...