Invasive Pressures
9650-0806-01 Rev. D
Propaq MD Operator’s Guide
A-23
Typical Determination time without Artifact:
Measurements on the deflation:
30 to 45 seconds
Measurements on the inflation (SureBP)*:
15 to 30 seconds
* using dual lumen cuffs
Maximum Determination Time - Measurement on the Inflation
Adult: 150 seconds
Pediatric:120 seconds
Neonatal:80 seconds
Blood Pressure Validation
:
Blood pressure measurements determined with this device are equivalent to those obtained by a
trained observer using the cuff/stethoscope auscultation method for adults and pediatric patients
and equivalent to an intra-arterial measurement for neonatal patients, within the limits
prescribed by the American National Standards Institute (ANSI-AAMI SP10). To receive a
copy of the report containing the AAMI SP10 results, contact the ZOLL Technical Service
Department.
Invasive Pressures
Number of Channels
: 3
Pressure range
: -30 to 300 mmHg
Pressure Accuracy
: +/- 2 mmHg or 2% of reading, whichever is greater, plus transducer error.
Pulse Rate Range:
25 to 250 BPM
Pulse Rate Accuracy:
+/- 3 BPM, or +/- 3% of value whichever is greater
Pulse Rate Display:
Average of last 4 beat-to-beat intervals.
Zero Adjust:
+ / - 200 mmHg
Transducer
:
Sensitivity: 5uV/V/mmHg
Offset: +/- 125 mmHg including transducer offset
Excitation Impedance Range: 150 to 10,000 ohms
Excitation Voltage: 4.75 +/- 0.25 VDC
Connector: 6-pin circular MS3100 series
Connect to:
A
B
C
D
E
Signal Type
Sig (-)
Exc (+)
Sig (+)
Exc (-)
shield
Summary of Contents for Propaq MD
Page 1: ...Propaq MD Operator s Guide 9650 0806 01 Rev D...
Page 40: ...CHAPTER 1 GENERAL INFORMATION 1 28 www zoll com 9650 0806 01 Rev D...
Page 60: ...CHAPTER 2 PRODUCT OVERVIEW 2 20 www zoll com 9650 0806 01 Rev D...
Page 74: ...CHAPTER 4 TRENDS 4 4 www zoll com 9650 0806 01 Rev D...
Page 84: ...CHAPTER 5 ALARMS 5 10 www zoll com 9650 0806 01 Rev D...
Page 94: ...CHAPTER 6 MONITORING ECG 6 10 www zoll com 9650 0806 01 Rev D...
Page 104: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 0806 01 Rev D...
Page 120: ...CHAPTER 8 MONITORING NON INVASIVE BLOOD PRESSURE NIBP 8 16 www zoll com 9650 0806 01 Rev D...
Page 154: ...CHAPTER 11 MONITORING INVASIVE PRESSURES IBP 11 10 www zoll com 9650 0806 01 Rev D...
Page 160: ...CHAPTER 12 MONITORING TEMPERATURE 12 6 www zoll com 9650 0806 01 Rev D...
Page 186: ...CHAPTER 14 12 LEAD ECG INTERPRETIVE ANALYSIS 14 16 www zoll com 9650 0806 01 Rev D...
Page 212: ...CHAPTER 17 ADVISORY CPR PROTOCOL DEFIBRILLATION 17 6 www zoll com 9650 0806 01 Rev D...
Page 222: ...CHAPTER 18 EXTERNAL PACING 18 10 www zoll com 9650 0806 01 Rev D...
Page 240: ...CHAPTER 21 PATIENT DATA 21 6 www zoll com 9650 0806 01 Rev D...
Page 268: ...CHAPTER 22 COMMUNICATIONS 22 28 www zoll com 9650 0806 01 Rev D...
Page 284: ...CHAPTER 24 MAINTENANCE 24 12 www zoll com 9650 0806 01 Rev D...
Page 326: ...ACCESSORIES B 6 www zoll com 9650 0806 01 Rev D...