Changing the Trends Status Window Display
9650-0806-01 Rev. D
Propaq MD Operator’s Guide
4-3
Changing the Trends Status Window Display
By default, the Trends status window displays all logged trend information. It displays the
numeric information for all monitoring functions, which the unit logs at a 30-second interval,
and when you take NIBP measurements, when a patient alarm occurs, and when you press
.
To configure the display of the Trends status window, press (
) and press the Trend Settings
quick access key (
) to display the Trends Settings control panel. On the Trends Settings
control panel, select
Trend Display Format
to specify the following monitored vital signs that
appear in the Trends status window:
Continuous Waveform Recording
Continuous waveform recording allows you to record continuous waveforms to a full
disclosure case log for the patient being monitored. This feature needs to be enabled by a
supervisor. The supervisor can access this feature by pressing the Setup quick access key (
),
and then selecting Supervisor>Log>Waveform Recording. In this menu, the following settings
can be customized:
•
Record Displayed Waveforms
– Record only the top waveform that is displayed or record
all waveforms that are displayed.
•
Record Additional Waveforms
– Always record the CO
2
waveform or always record the
IBP and the CO
2
waveforms.
There is an additional waveform that can be enabled in the Waveform Recording menu. This
option,
Record Pads Imp. Waveform
, measures the patient’s impedance between
defibrillation electrodes.
Note:
When Continuous waveform recording is disabled, no waveforms (except snapshots)
are recorded.
The full disclosure case can, at a minimum, concurrently store the following information:
•
32 monitor snapshots
•
500 non-ECG events
•
24 hours of continuous ECG (4 waveforms), Capnography, IBP (3 channels), and Pads
Impedance
The actual information that is stored depends on usage. Also, the specific combination of stored
continuous waveform data depends on how the waveform recording settings are configured in
the Supervisor menu.
Trend Format
Vital Signs Displayed
Resp
HR, SpO2, RR, EtCO2, FiCO2
NIBP
HR, SpO2, NIBP, RR
IBP1
HR, SpO2, IBP1, RR
IBP2
HR, SpO2, IBP2 RR
IBP3
HR, SpO2, IBP3, RR
Temp
HR, SpO2, T1, T2,
T
LOG
Summary of Contents for Propaq MD
Page 1: ...Propaq MD Operator s Guide 9650 0806 01 Rev D...
Page 40: ...CHAPTER 1 GENERAL INFORMATION 1 28 www zoll com 9650 0806 01 Rev D...
Page 60: ...CHAPTER 2 PRODUCT OVERVIEW 2 20 www zoll com 9650 0806 01 Rev D...
Page 74: ...CHAPTER 4 TRENDS 4 4 www zoll com 9650 0806 01 Rev D...
Page 84: ...CHAPTER 5 ALARMS 5 10 www zoll com 9650 0806 01 Rev D...
Page 94: ...CHAPTER 6 MONITORING ECG 6 10 www zoll com 9650 0806 01 Rev D...
Page 104: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 0806 01 Rev D...
Page 120: ...CHAPTER 8 MONITORING NON INVASIVE BLOOD PRESSURE NIBP 8 16 www zoll com 9650 0806 01 Rev D...
Page 154: ...CHAPTER 11 MONITORING INVASIVE PRESSURES IBP 11 10 www zoll com 9650 0806 01 Rev D...
Page 160: ...CHAPTER 12 MONITORING TEMPERATURE 12 6 www zoll com 9650 0806 01 Rev D...
Page 186: ...CHAPTER 14 12 LEAD ECG INTERPRETIVE ANALYSIS 14 16 www zoll com 9650 0806 01 Rev D...
Page 212: ...CHAPTER 17 ADVISORY CPR PROTOCOL DEFIBRILLATION 17 6 www zoll com 9650 0806 01 Rev D...
Page 222: ...CHAPTER 18 EXTERNAL PACING 18 10 www zoll com 9650 0806 01 Rev D...
Page 240: ...CHAPTER 21 PATIENT DATA 21 6 www zoll com 9650 0806 01 Rev D...
Page 268: ...CHAPTER 22 COMMUNICATIONS 22 28 www zoll com 9650 0806 01 Rev D...
Page 284: ...CHAPTER 24 MAINTENANCE 24 12 www zoll com 9650 0806 01 Rev D...
Page 326: ...ACCESSORIES B 6 www zoll com 9650 0806 01 Rev D...