C
HAPTER
6 M
ONITORING
ECG
6-4
www.zoll.com
9650-0806-01 Rev. D
5-Lead Electrode Placement
Depending upon local usage, the ECG leads are marked either RA, LA, LL, RL, and V or R, L,
F, N and C. The following table shows the markings and color codes for the different lead sets.
AHA Color Coding
IEC Color Coding
Placement of Electrodes
RA/White Electrode
R/Red Electrode
Place near patient’s right mid-clavicular line,
directly below clavicle.
LA/Black Electrode
L/Yellow Electrode
Place near patient’s left mid-clavicular line,
directly below clavicle.
LL/Red Electrode
F/Green Electrode
Place between 6th and 7th intercostal
space on patient’s left mid-clavicular line.
RL/Green* Electrode
N/Black* Electrode
Place between 6th and 7th intercostal
space on patient’s right mid-clavicular line.
V/Brown* Electrode
C/White* Electrode
Single movable chest electrode. Place this
electrode in one of the positions, V1 - V6,
as shown in the following figure.
V1 -- 4th intercostal space at right sternal
margin.
V2 -- 4th intercostal space at left sternal
margin.
V3 -- Midway between V2 and V4 leads.
V4 -- 5th intercostal space at mid-clavicular
line.
V5 -- Same transverse level as V4 at left
anterior-axillary line.
V6 -- Same transverse level as V4 at left
mid-axillary line.
Summary of Contents for Propaq MD
Page 1: ...Propaq MD Operator s Guide 9650 0806 01 Rev D...
Page 40: ...CHAPTER 1 GENERAL INFORMATION 1 28 www zoll com 9650 0806 01 Rev D...
Page 60: ...CHAPTER 2 PRODUCT OVERVIEW 2 20 www zoll com 9650 0806 01 Rev D...
Page 74: ...CHAPTER 4 TRENDS 4 4 www zoll com 9650 0806 01 Rev D...
Page 84: ...CHAPTER 5 ALARMS 5 10 www zoll com 9650 0806 01 Rev D...
Page 94: ...CHAPTER 6 MONITORING ECG 6 10 www zoll com 9650 0806 01 Rev D...
Page 104: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 0806 01 Rev D...
Page 120: ...CHAPTER 8 MONITORING NON INVASIVE BLOOD PRESSURE NIBP 8 16 www zoll com 9650 0806 01 Rev D...
Page 154: ...CHAPTER 11 MONITORING INVASIVE PRESSURES IBP 11 10 www zoll com 9650 0806 01 Rev D...
Page 160: ...CHAPTER 12 MONITORING TEMPERATURE 12 6 www zoll com 9650 0806 01 Rev D...
Page 186: ...CHAPTER 14 12 LEAD ECG INTERPRETIVE ANALYSIS 14 16 www zoll com 9650 0806 01 Rev D...
Page 212: ...CHAPTER 17 ADVISORY CPR PROTOCOL DEFIBRILLATION 17 6 www zoll com 9650 0806 01 Rev D...
Page 222: ...CHAPTER 18 EXTERNAL PACING 18 10 www zoll com 9650 0806 01 Rev D...
Page 240: ...CHAPTER 21 PATIENT DATA 21 6 www zoll com 9650 0806 01 Rev D...
Page 268: ...CHAPTER 22 COMMUNICATIONS 22 28 www zoll com 9650 0806 01 Rev D...
Page 284: ...CHAPTER 24 MAINTENANCE 24 12 www zoll com 9650 0806 01 Rev D...
Page 326: ...ACCESSORIES B 6 www zoll com 9650 0806 01 Rev D...