Chapter 7 Monitoring Respiration (Resp) and Heart Rate (HR)
7-4
www.zoll.com
9650-0806-01 Rev. D
Respiration Rate Alarm Limits
Initially, the
Resp Alarm
Settings
menu specifies that Resp
alarms are enabled (
ON
)
or
disabled (
OFF
),
and displays the default Upper and Lower respiration rate alarm limits. The
upper and lower limits can be
ON
or
OFF
(default is
OFF
).The following table lists the default
respiration rate alarm limits for adult, pediatric, and neonate patients, and gives the range in
which you can set these limits:
Using the Resp Parameter Control Panel
To configure alarms through the
Resp
Parameter Control Panel, use the navigation keys to
highlight and select the Respiration Rate meter and display the
Resp
Parameter Control Panel:
Figure 7-1 Respiration Parameter Control Panel
The
Resp
Parameter Control Panel allows you to set the following parameters:
•
RR/BR Alarm
-- enable/disable the Resp alarm and set high/low alarm limits.
•
No Breath Alarm
-- sets the duration of the No Breath alarm, or disables alarm by selecting
“Off”.
•
Resp Monitoring --
enable/disable respiration monitoring.
•
Resp Lead
-- selects the Resp lead, Lead I (RA-LA) or Lead II (RA-LL), from which the
Propaq MD unit calculates the respiration rate. Resp Lead selection is independent of ECG
Lead selection.
•
CO2/Resp Sweep Speed --
sets the respiratory sweep speed on the display.
Patient Type
Respiration Rate Default
Respiration Rate Range
Adult
Lower: 3 br/min
Upper: 50 br/min
Lower: 0 to 145 br/min
Upper: 5 to 150 br/min
Pediatric
Lower: 38 br/min
Upper: 50 br/min
Lower: 0 to 145 br/min
Upper: 5 to 150 br/min
Neonate
Lower: 12 br/min
Upper: 80 br/min
Lower: 0 to 145 br/min
Upper: 5 to 150 br/min
Resp
Lower
Upper
RR/BR Alarm
10
30
No Breath Alarm
30 sec
Resp Monitoring
On
Resp Lead
Lead II (RA-LL)
CO2/Resp Sweep Speed
12.5 mm/s
Summary of Contents for Propaq MD
Page 1: ...Propaq MD Operator s Guide 9650 0806 01 Rev D...
Page 40: ...CHAPTER 1 GENERAL INFORMATION 1 28 www zoll com 9650 0806 01 Rev D...
Page 60: ...CHAPTER 2 PRODUCT OVERVIEW 2 20 www zoll com 9650 0806 01 Rev D...
Page 74: ...CHAPTER 4 TRENDS 4 4 www zoll com 9650 0806 01 Rev D...
Page 84: ...CHAPTER 5 ALARMS 5 10 www zoll com 9650 0806 01 Rev D...
Page 94: ...CHAPTER 6 MONITORING ECG 6 10 www zoll com 9650 0806 01 Rev D...
Page 104: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 0806 01 Rev D...
Page 120: ...CHAPTER 8 MONITORING NON INVASIVE BLOOD PRESSURE NIBP 8 16 www zoll com 9650 0806 01 Rev D...
Page 154: ...CHAPTER 11 MONITORING INVASIVE PRESSURES IBP 11 10 www zoll com 9650 0806 01 Rev D...
Page 160: ...CHAPTER 12 MONITORING TEMPERATURE 12 6 www zoll com 9650 0806 01 Rev D...
Page 186: ...CHAPTER 14 12 LEAD ECG INTERPRETIVE ANALYSIS 14 16 www zoll com 9650 0806 01 Rev D...
Page 212: ...CHAPTER 17 ADVISORY CPR PROTOCOL DEFIBRILLATION 17 6 www zoll com 9650 0806 01 Rev D...
Page 222: ...CHAPTER 18 EXTERNAL PACING 18 10 www zoll com 9650 0806 01 Rev D...
Page 240: ...CHAPTER 21 PATIENT DATA 21 6 www zoll com 9650 0806 01 Rev D...
Page 268: ...CHAPTER 22 COMMUNICATIONS 22 28 www zoll com 9650 0806 01 Rev D...
Page 284: ...CHAPTER 24 MAINTENANCE 24 12 www zoll com 9650 0806 01 Rev D...
Page 326: ...ACCESSORIES B 6 www zoll com 9650 0806 01 Rev D...