Quick Reference Guide

1. Press        to bring up the Pacer menu.
2. Confirm Pacer Mode and Pacer Rate.
3. Press        &        to highlight 

Start Pacer

.

4. Press         to begin pacing.
5. Press        &        to highlight the Pacer Output. Press       .
6. Press        &        to adjust the pacer output until capture occurs, 
      then decrease the output to the lowest level that still maintains
      capture.
7. Press         to exit the Pacer menu.

Starting the Pacer & Changing Pacer Settings

1. Press        to bring up the Pacer menu.
2. Press        &        to highlight 

Pause Pacer

.

3. Press         to pause the pacer. 

Pausing the Pacer

1. Press        to bring up the Pacer menu.
2. Press        &        to highlight 

Turn Pacer Off

.

3. Press         to turn the pacer off.

Turning Off the Pacer

Pacer

Pacer

Pacer

All manuals and user guides at all-guides.com

Summary of Contents for Propaq MD

Page 1: ...P Start Stop Ready For Use Indicator 3 Shock 2 Charge 1 Select Energy Sync Pacer Battery Charge LED Auxiliary Power LED Data Transfer LED Equipment Alert LED Patient Alarm LED Power Button Align arrow...

Page 2: ...ns Changing ScreenViews Press to cycle through the different screen brightness options Changing Display Brightness Taking a Monitor Snapshot Press to take a monitor snapshot The unit records all wavef...

Page 3: ...ce Press Changing the Lead Source inWaveformTrace 1 4 1 Press to highlight the Gain Label above the waveform Press 2 Press to highlight the desired waveform size Press Changing theWaveform Gain 5 Lead...

Page 4: ...shot 1 Press to display the 12 Lead Review menu 2 Press to highlight the desired 12 lead snapshot Press Reviewing a Stored 12 Lead Snapshot Changing Displayed Leads in a 12 Lead Snapshot Press to prin...

Page 5: ...ess to initiate a manual NIBP reading Initiating an NIBP Reading 1 Press to highlight the NIBP numeric Press 2 Press to highlight NIBP Mode Press 3 Press to toggle to Auto Mode Press Setting Up Automa...

Page 6: ...eth waveform Press 2 Press to highlight the desired waveform size Press Changing theWaveform Gain 1 Press to highlight the Source Label above the pleth waveform Press 2 Press to highlight Remove Press...

Page 7: ...hen CO2 is on and the unit automatically displays the capnogram and the EtCO2 value Starting CO2 Monitoring 1 Press to highlight the Scale Label above the CO2 waveform Press 2 Press to highlight the d...

Page 8: ...r Limit Press 6 Press to adjust the setting to the desired value e g 120 Press Note An alarm must be enabled turned on before the alarm limit can be adjusted Enabling Individual Alarms Setting Alarm L...

Page 9: ...orming Defibrillation While the Propaq MD unit is charging or charged press to safely disarm the energy Changing the selected energy will also disarm the Propaq MD unit Disarming the Defibrillator 1 P...

Page 10: ...then decrease the output to the lowest level that still maintains capture 7 Press to exit the Pacer menu Starting the Pacer Changing Pacer Settings 1 Press to bring up the Pacer menu 2 Press to highl...

Reviews: