Advisory Defibrillation Procedure
9650-0912-06 Rev. G
ZOLL R Series Operator’s Guide
4–5
•
Units with the automatic charge option disabled will alternately display the messages
SHOCK ADVISED
and
PRESS CHARGE.
Press the
CHARGE
button.
Regardless of the analysis result, the user can control the defibrillator manually. For example,
the user can defibrillate the patient even if the advisory function issues a NO SHOCK ADV.
message.
3 Press SHOCK
WARNING!
Warn all persons in attendance of the patient to STAND CLEAR prior to defibrillator
discharge.
Do not touch the bed, patient, or any equipment connected to the patient during
defibrillation. A severe shock can result. Do not allow exposed portions of the patient’s
body to come in contact with metal objects, such as a bed frame, as unwanted pathways
for defibrillation current may result.
DEFIB 120J SEL.
Code
Marker
Param
Sync
On/Off
00:01
SHOCK ADVISED
ECG
Options
Alarms
Report
Data