C
HAPTER
5
S
YNCHRONIZED
C
ARDIOVERSION
5–8
www.zoll.com
9650-0912-06 Rev. G
Once the energy is delivered, the display simultaneously shows
XXXJ DELIVERED
and
DEFIB
XXXJ SEL
. After approximately 5 seconds, the
XXXJ DELIVERED
message disappears and the
DEFIB XXXJ SEL
. message remains to indicate the selected energy level.
If additional countershocks are necessary, readjust the energy level as necessary, press the
Sync On/Off
, and then the
Remote Sync
softkeys and repeat. Note that
REMOTE SYNC XXXJ
SEL
must be displayed prior to pressing the
CHARGE
button.
If the
ANALYZE
button is pressed while the unit is in Remote Sync mode, the unit displays the
REMOVE SYNC
message and disallows ECG rhythm analysis until the unit is taken out of Sync
mode.