R Series ALS In-service Guide

 

 

OVERVIEW

 

 

Color Coded Operating System

  Review with the Mode Selector 

dial. 

o

 

RED 

for Defibrillation 

o

 

GREEN 

for Pacing 

o

 

GREY 

for Monitoring. 

 

Power Indicators 

 

Battery Indicator 

o

 

GREEN 

= Charged 

o

 

AMBER 

= Charging 

o

  FLASHING 

= Problem (check battery) 

o

 

Indicator lights on battery show 30 min. increments 
for a total of 4 hours of runtime. 

o

 

If orange light appears on “?”of battery, call BioMed 
to reset. 

 

A/C Power Indicator 

o

 

GREEN 

= Device is plugged in 

o

  No light = Device is NOT plugged in 

 

Test & Code Readiness Indicator 

  Every 24 hours, the device will do a 100 point test and will display a 

   

Green check in window = Device is ready for use 

   

Red x in window = Failed a part of the test.  Device is NOT ready for use. 

  The device will display “Readiness Test Failed” on the screen and will 

highlight what needs to be fixed. These are usually easily fixable problems 
(No pads plugged in, no battery, no A/C power, etc). 

  After fixing any easily fixable problems, perform a 

manual test 

at 30J with 

pads plugged in. “30J Test OK.” “Defib Pad Short.” and green check in 
window indicates pass. 

  If the device reads “DEFIB FAILURE”, call BioMed. 

 

Setup 

  A OneStep CPR Electrode should 

ALWAYS 

be connected to the therapy cable. 

o

  Ensures “Code-Readiness” 

o

  Allows the device to pass its Automated Daily Test. 

 

Checks 

 

Daily Check:  

2 lights, a green check mark, the cap snugly on the back 

 

Weekly Check: 

Follow the “Weekly Test” instructions in your Code Log books 

 

Recorder Tray 

  Press button to open door 

  Insert the paper with the black arrow facing towards the back of the device 

  Pull a strip or two forward and let paper drop 

  Close the door on top of the strip. 