See-Thru CPR

 

 

Put rhythm s

imulator on “

VF Artifact”

 

 

Two Wave Forms 

 

Raw EKG 

- Waveform with artifact 

 

Filtered EKG 

Internal algorithm filters out artifact allowing you to see the underlying rhythm. 

o

  You can see the underlying rhythm without stopping compressions. 

o

  Someone can press charge while you continue doing compressions. 

o

  Once the red Charge button lights up and you hear the tone… 

  Stop compressions and evaluate the rhythm to make sure it is 

shockable 

  Then, shock and get back on the chest in less than 10 seconds to 

meet guidelines 

 

PACING (

Green Zone

 

Put rhythm s

imulator on “3rd”

 

Leads 

  Built into OneStep Complete and Pacing Pads 

  Apply 3 lead if you are not using OnStep or Pacing Pads. 

  Apply 3 lead if you use alternate placement of OneStep Pads 

Dials 

 

1.  Rate 
2.  Output 

Default Settings 

  0mA and 70PPM

 

Output 

  40-80mA. (Good starting point)

 

Pacing 

  Confirmed with downward pacing markers

 

Capture 

  Widened QRS after each pacer marker

 

  Set output 10% above threshold

 

  Confirm by checking patient pulse

 

Underlying Rhythm 

  Press and hold the 4:1 button at the button of the screen to see underlying rhythm

 

 

CARDIOVERSION (

Red Zone

)

 

 

Put rhythm s

imulator on “AFIB”

 

 

1. Turn to 

DEFIB 

mode 

2. Press 

Sync On/Off 

button on the bottom right 

(soft key) 

3. Confirm arrow markers on every “R” wave telling 

the defib where to shock. 

4. 

Press 

CHARGE 

5. 

“All clear!”, 

then press and hold 

SHOCK 

until 

the device fires. 