Specifications
A-2
www.zoll.com
9650-002360-01 Rev. A
Oxygen Input Pressure
Nominal: 55 psig (380 kPa)
Range: 40 to 87 psig (280 to 600 kPa)
Mean Airway Pressure
(MAP)
Reading: 0 to 99.9 cm H
2
O +/- (2 cm H
2
O + 8% of reading)
Resolution: 1 cm H
2
O
Airway Pressure High
Limit
Setting: 20 to 100 cm H
2
O
Setting Resolution: 1 cm H
2
O
Airway Pressure Low
Limit
Setting: Off, 3 to 35 cm H
2
O
Setting Resolution: 1 cm H
2
O
Breath Trigger
-6.0 to -0.5 cm H
2
O +/- (0.25 cm H
2
O + 5% of setting below)
Airway Pressure
Waveform
0 to 99 cm H
2
O +/- (2 cm H
2
O + 8% of reading)
Minute Volume
0 to 99.9 lpm +/- (0.1 lpm + 8% of reading)
LED Status/Alarm
Indicator
Red, Yellow, and Green
Alarm Volume
82 dBA @ 1 meter
Noise Level
~60 dBA when measured at 1 meter (operating at default settings using the
compressor only)
Operating Voltages
AC Supply: 100 to 240 VAC (50/60 and 400 Hz) -- use only the AC/DC power supply
that ZOLL provides with the device.
DC Supply: Nominal 12.5 to 28.0 VDC (accepts DC voltages between
11.8 to 30 VDC).
Operating Time
Internal Battery
10 hours at default settings
Ventilator Temperature
Ranges
Standard Operating Temperature: -10
C to 40
C (14
F to 104
F)
Extended Operating Temperature: -13
C to 49
C (9
F to 120
F)
Extreme Operating Temperature: -26
C to 55
C (13
F to 131
F)
Battery Temperature
Ranges
Battery Charge: 0
C to 45
C (32
F to 113
F)
Battery Discharge: up to 75
C (167
F)
Size
8.0” Wide X 12.5” High X 4.5” Deep (20.3 cm Wide X 31.8 cm High X 11.4 cm Deep)
Weight
~9.7 lbs (4.4 kg)
Warranty
Limited, 1 year
Altitude
-685.8 m (-2,250 ft) to 7620 m (25,000 ft) above sea level or 110 to 37.6 kPa
Operational Humidity
15% to 95% non-condensing
Transport and Storage
Temperature: EN1789 for ambulance 6.3.2.1 Storage Temperature test
Humidity: 15 to 95 % RH (non-condensing)
Vibration
IEC 60068-2-6, IEC 60068-2-34, IEC 60068-2-36, IEC 60068-2-64
Shock
IEC 60068-2-27
Bump
IEC 60068-2-29
Shock/Vibration
MIL SRD 810G, Method 514.6, Procedure I:Jet Aircraft, Fixed Wing, and Rotary Wing
Parameter
Operating Range