Specifications
A-4
www.zoll.com
9650-002360-01 Rev. A
The device meets the electromagnetic tests as specified by regulations. The following tables
provide guidance as to the environments in which you can operate the device.
The ventilator is intended for use in the electromagnetic environment specified below. The
customer or user of the ventilator should ensure that they are used in such an environment.
Method of Sterilization or
Disinfection
O
2
Supply hoses and connections should be wiped with a
damp, soapy cloth and thoroughly dried with a
lint-free cloth. The device’s housing should also be cleaned as
necessary with a damp, soapy cloth and throughly dried with a
lint-free cloth. Do not clean with abrasives or chlorinated
hydrocarbon cleansers.
Ventilator circuits are only for single use. Follow all IFU
instructions.
Degree of Safety of Application in
the presence of a Flammable
Anesthetic Mixture with Air or with
Oxygen or Nitrous Oxide
Equipment
not
suitable for use in presence of Flammable
Anesthetic Mixture of Air or with Oxygen or Nitrous Oxide
Mode of Operation
Continuous Operation
Category
Classification
Emissions test
Compliance
Electromagnetic environment-guidance
RF emissions
CISPR 11
Group 1
The ventilator used RF energy only for their internal function.
Therefore, their RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class A
The ventilator is suitable for use in all establishments other than
domestic and those directly connected to the public
low-voltage power supply network that supplies buildings used
for domestic purposes.
Harmonic emissions
IEF 61000-3-2
Class A
Voltage
fluctuations/flicker
emissions
IEC 61000-3-3
Complies
Immunity Test
Test level
Compliance level
Electromagnetic environment-
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
+/- 6 kV contact
+/- 8 kV air
+/- 6 kV contact
+/- 8 kV air
Floors should be wood, concrete, or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
+/- 2 kV for power
supply lines
+/- 1 kV for
input/output lines
+/- 2 kV for power
supply lines
+/- 1 kV for
input/output lines
Mains power quality should be that of a
typical commercial or hospital
environment.
Surge
IEC 61000-4-5
+/- 1 kV line(s) to
line(s)
+/- 2 kV line(s) to earth
+/- 1 kV line(s) to
line(s)
+/- 2 kV line(s) to earth
Mains power quality should be that of a
typical commercial or hospital
environment.