Alarms
5-16
www.zoll.com
9650-002360-01 Rev. A
Alarm Summary
A list of all ventilator alarms follows and divided into three categories:
•
High Priority
•
Medium Priority
•
Low Priority
Alarms in each list are provided in numerical order based on their Service Code.
High Priority Alarms
Backup
Apnea
PD Mode
AC(P)
SIMV(P)
AC(P)
BPM
12
12
14
I:E
1:4
1:3
1:3
PIP
20
20
20
PEEP
5
As Set
5
Rise Time
5
As Set
6
PIP High
25
As Set
25
PIP Low
3
As Set
10
Vt High
700
As Set
750
Vt Low
As Set
250
FIO
2
As Set
As Set
21
Trigger
-2 cm H
2
O
As Set
-2 cm H
2
O
Service Code
Alarm Name/Mitigation/Resolution
1001
Self Check Failure - Pneumatic System: Compressor
Alarm triggers when the compressor fails to operate or fails to provide the flow
required to deliver a breath and high pressure O
2
is not available to provide
ventilation.
Mitigation/Info: Pneumatic System: Compressor, Manually Ventilate
Patient, Connect 55 psig/380 kPa O
2
, Restart Ventilator with O
2
,
**Contact Service Center**