Product Overview
2-8
www.zoll.com
9650-002360-01 Rev. A
Pulse Oximeter Compatibility
The ventilator can accommodate an optional connection of external Masimo Pulse Oximeter.
When the appropriate sensor is connected, the pulse oximeter provides continuous noninvasive
monitoring of the oxyhemoglobin saturation (SpO
2
) and pulse rate (measured by the SpO
2
sensor) for adult, pediatric and infant patients.
The Masimo LNCS series of probes are approved for use with the ventilator. The Accessory
table in Appendix A lists the sensors which are available for use with the ventilator.
Power Sources
The ventilator can operate using external power or it can operate powered by its internal Li-ion
battery.
The external AC/DC Power cable is a universal supply that can operate with an input of
100 to 240 VAC 50/60 Hz. The external supply can also power the device when provided with
a 400 Hz input.
The external AC/DC Power cable that ZOLL provides with the ventilator delivers a DC input to
the device of 24 V at 4.2 A. When this external power source is present, the ventilator
automatically charges its internal battery while operating.
Only use the external power supply provided with the ventilator when connecting to AC power.
This power supply provides both Class I and Class II protection.
Operating Using External DC Power
The ventilator can also operate using external DC power. When connected to a standard vehicle
DC outlet using either the 12 or 28 VDC Power Cable that ZOLL offers, the ventilator
automatically charges its internal battery while operating.
Note:
The input connector of the ventilator accepts DC voltages between 11.8 to 30.0 VDC.
Caution
When using the standard vehicle DC outlet, do not jump start the vehicle during operation of the
ventilator.
Operating Using Battery Power
When an external power failure occurs, the ventilator automatically switches to its internal
battery for operating power and activates the EXTERNAL POWER FAILURE alarm; there is
no interruption in operation. When external power returns, operating power automatically
switches to the external power source and the following symbol displays on the ventilator
screen as shown in Figure 2-7.