Using the Z Vent Ventilator
9650-002360-01 Rev. A
ZOLL Z Vent Ventilator Operator’s Guide
4-3
Pediatric Default Parameter Setting Values
Mask CPAP Default Parameter Setting Values
Making a Choice From the Start Menu
Choose the option that is most is appropriate for the patient. With the Start Menu displayed,
follow these steps:
1. Turn the
Dial
to highlight your choice. For example, to choose the Mask CPAP default, turn
the
Dial
until Mask CPAP highlights.
2.
Press the
Accept
button to enable your choice. The ventilator begins operation with the
choice you made. For initial set-up, see Chapter 3.
Note:
The ventilator provides a 120-second alarm mute it automatically clears as described
above to allow the provider time to adjust parameter settings for the patient.
Parameter
Default Setting Values
Mode
SIMV (P)*
AC(P)
BPM
20
20
Ti
0.6
0.6
PIP
20
20
PEEP
4
4
PIP Limit
30
30
FIO
2
21
21
* Note: SIMV (V) and SIMV (P) are optional modes on the Z Vent.
Parameter
Default Setting Value
Mode
CPAP
Backup BPM
12
Backup I:E
1:3
Backup PIP
20
PEEP
5
PIP Limit
30
FIO
2
21