16
Seite
Seite
Seite
Seite
Seite
Page
1.3.6 Diagnostic tests
❶
Run pre-heat/dry program
This program is run prior to the first operation of the
furnace or when the furnace has been stored or
turned off for long durations.(see 0.5 Set-up oven)
Note: The lift remains in a slightly open position, so
any existing residual moisture can easily escape.
❷
Run purge program
Use this high temperature program to remove
contaminates and impurities.
Never use ANY purging additives such as carbon
(graphite pellets) for cleaning the firing chamber!
❸
About version
Displays the software version currently installed.
❹
Update software
Requires the entry of a password.
(We will provide the password and instructions with
software updates.)
❺
Service tests
Vacuum test: determines the available vacuum for
your location and performs a leakage test.
Factory tests: for use by service center.
1.3.7 Customize country
This section sets the language as well as the date,
time and units (metric/imperial).
1.3.8 Adjust display
The brightness of the display is set using this functi-
on (recommended value 95).
1.3.9 Audio Beeps
Software switch to allow beep or not
1.3.10 Set preview mode
This function displays the parameters of the selec-
ted program before starting it.
1.3.11 Initial oven setup
see 0.5
1 Operation
Page