Ref.No.
:
082019
7. rcFSMode:
If user set or calibrated the sensor wrongly, you can recover the factory setting
7.1. Press the Mode button (
○
5
), until rcFS flashes on the lower display
7.2. Press Enter (
○
2
), “no” shows on the upper display
7.3. Press Up/Down (
○
4
/
○
3
), “no” change to “yes”
Press Enter (
○
2
), Factory Setting will reload, return to UserMode;
7.4. Press Mode (
○
5
), leave rcFS Mode, return to UserMode
SPECIFICATIONS (Preliminary Version)
Method
- Dual Beam (NDIR Non-Dispersive-Infrared)
Display
- LCD
Independent CO
2
and Temperature readings.
Calculates and Displays Ventilation Rates
Sample Method -
Diffusion or flow through (50 ~200 ml/min)
■
Performance - CO
2
Measurement Range
0-3,000 ppm display
Display Resolution
1ppm at 0~1,000ppm; 5ppm at 1,001~2,000ppm; 10ppm at 2,001~3,000ppm
Accuracy
±50 ppm or ±5% of reading
Repeatability
±20 ppm
Temperature Dependence
Typ.
±0.1% of reading per °C or ±2 ppm per °C, whichever is greater, referenced to 25°C
Pressure Dependence
0.13% of reading per mm Hg (Corrected via user input for altitude)
Response Time
About 2min for 63% of step change
Warm-Up Time
<60 seconds at 22°C
■
Performance - Temperature
Temperature Range
Display 32 to 122°F (0 to 50°C)
Display Resolution
0.1°F (0.1°C)
Display Options
°F / °C, or Off. Set with Up / Down button
Accuracy
±2°F (±1°C)
When the fan blows to the device directly, the accuracy of temperature is + / -1.5 degC .
Response Time
20-30 minutes (case must equilibrate with environment)
■
Outputs
OC:
Normally Low, 100mA max. @ 24 VDC. Adjustable setpoint, factory setting is 1000ppm, hysteresis is 50ppm
■
Power Supply
Two power supply types
Battery type: Alkaline, AA* 4 (80 hours)
External: 6 VDC from external AC/DC adapter which is included in package (Use specified AC adapter only)
Power Requiremen
t
160 mA Peak, 15 mA average from 6V
CAUTION
The device can be supplied either by batteries or an AC adapter.
DO NOT
install batteries and plug in an AC adapter at the same time.
■
General Operating Conditions
Operating Temperature:
32-122°F (0-50°C) 0-95% RH, non-condensing
Storage Temperature:
-4 to 140°F (-20 to 60°C)
■
Dimensions
:
24.3×81.1×165 mm (0.96×3.19×6.50 inch)
Weight (including batteries)
:
273.6 grams (9.65 oz)
■
LCD ERROR MESSAGES
The device incorporates visual diagnostic messages as follows:
「
」
Error 3
, it
is displayed when the temperature being measured is outside of the measurement range.
「
」
「
」
Error 5~6,
it is necessary to reset the device. To reset it, turn the instrument off, remove the battery and wait for a
minimum of one minute, reinsert the battery and turn on. If the error message remains please contact the Service
Department for further assistance.