 Attest

 290G Auto-reader

Reference Tool

Refer to 3M

 Attest

 290G Auto-reader Operator’s Manual for complete instructions 

Refer to 3M

 Attest

 Rapid Readout Biological Indicator (BI) package inserts for complete instructions 

Model 290G Auto-reader for use with:

3M

 Attest

 1294 Rapid Readout  

Biological Indicators for Ethylene Oxide

Caution Code

What does this mean?

Solution

C1

Unit Warm-up Caution

• Incubation block not at temperature
• BI readings are disabled until block reaches 37˚C±2˚C

Allow 30 minute warm-up after Attest Auto-reader 

 

is plugged in

C2

 BI has been removed from an incubation well before 
incubation is complete

  

•  All 3 lights for affected BI will light 

 

and alarm will sound, if active

Replace BI back into the same incubation well within 

 

10 seconds

Error Code

What does this mean?

Solution

E1

Temperature Control Error

•  System no longer able to control temp of incubator block

Unplug Attest Auto-reader and plug back 

 

in to initiate the self-diagnostic test*

E3

Memory Error

Unplug Attest Auto-reader and plug back 

 

in to initiate the self-diagnostic test*

E4

Analog-digital Conversion Error

Unplug Attest Auto-reader and plug back 

 

in to initiate the self-diagnostic test*

E7

Communication Error

•  Internal communications failed

Unplug Attest Auto-reader and plug back 

 

in to initiate the self-diagnostic test*

E8

Bulb Error

• System detects the bulb is too dim to properly read BI results

Unplug Attest Auto-reader and plug back 

 

in to initiate the self-diagnostic test*

E9

Transport Error

•  System detects motor failure or positioning problem

Unplug Attest Auto-reader and plug back 

 

in to initiate the self-diagnostic test*

  *

 If after initiating self-diagnostic test the error code recurs, call for assistance (see phone number on back); unit may need repair/replacement.

Function setting

**

What does this mean?

Fd

Attest Auto-reader is configured to read:

Attest 1294 BI in wells 1–12, 4-hour incubation. Corresponding well label should be green.

**  Configuration Set-Up:

 Attest Auto-reader is configured for the following configuration function setting. 

 

When unit is powered on, Fd will momentarily appear in the “digital display window” on front of unit.