background image

Charge Analyzer 711

Operating Instructions

August 2005

78-8136-3182-3

Distributed by: 

 

All-Spec Industries 

5228 US Hwy 421 N 

Wilmington, NC 28401 

sales@all-spec.com 

www.all-spec.com 

PH: 800-537-0351 

FX: 800-379-9903 

 

Summary of Contents for Charge Analyzer 711

Page 1: ...e Analyzer 711 Operating Instructions August 2005 78 8136 3182 3 Distributed by All Spec Industries 5228 US Hwy 421 N Wilmington NC 28401 sales all spec com www all spec com PH 800 537 0351 FX 800 379...

Page 2: ...6 3 Specifications 7 4 Fieldmeter Operation 8 5 Voltmeter Operation 10 6 Static Decay Time Balance Operation 13 7 Remote Field Sensor Probe Operation 18 8 Data Recording Analog Outputs 20 9 Verificat...

Page 3: ...ry and or property damage WARNING Indicates a potentially hazardous situation which if not avoided could result in death or serious injury and or property damage CAUTION Indicates a potentially hazard...

Page 4: ...at contains lead in the solder At the end of service life dispose of the charge analyzer in accordance with federal state and local requirements CAUTION To reduce the risks associated with property da...

Page 5: ...of the analyzer can be maintained through the AC adaptor Recharge time is approximately 14 hours with unit off when batteries are fully discharged Important Note Avoid extreme discharge of the rechar...

Page 6: ...red cover F Teflon measuring lead 1m G RS 232 interface cable H 50 mm 2 inches high conductive con tainer with insulating handle I Insulated bulldog clip J Metal spacers 3 with thread 76 mm 3 in leng...

Page 7: ...ment Ranges Distance Voltage Resolution 1 cm 0 to 10kv 1V 2 5 cm 0 to 20kv 2V 5 cm 0 to 50kv 10V 10 cm 0 to 100kv 10V 20 cm 0 to 200kv 20V Displays Two 11 segment positive negative LED bar charge indi...

Page 8: ...amaged Do not attempt to modify or repair no user serviceable parts inside contact 3M Service for repair To reduce the risks associated with fire or explosion Do not operate the analyzer in an explosi...

Page 9: ...r The digital display reads the field strength in kV m To convert readings of kV m into volts use the following formulas R 100 x D cm or R 39 37 x D in where R Reading indicated on the display and D D...

Page 10: ...zer 711 in Voltmeter mode upper left corner R Voltage measurement range selected If range selected is too low OVERFLOW will appear AR Auto voltage measurement range selected arrow symbol will briefly...

Page 11: ...l of the unit to an earth electrical ground 6 Depress the on off button to activate the analyzer The display will briefly indicate CHANGE MODE OF OPERATION then VOLTMETER IS ENABLED if cup has not bee...

Page 12: ...on a person Demonstration 3 Performance of an ESD Control Flooring 1 Stand on a non static control flooring e g mat tile epoxy etc 2 Hold the cylinder electrode in one hand and perform normal body mo...

Page 13: ...ESD control surface Demonstration 5 Discharge Performance of a Static Control Packaging Material 1 Place a packaging material on top of a grounded static control worksurface 2 Charge up the conductive...

Page 14: ...ge selection to perform the zeroing This completes the zeroing of all ranges Setting the START STOP voltage Note The Start voltage can only be adjusted from 600 to 1100 volts The Stop voltage can only...

Page 15: ...ime period The unit will not have an active timer in this mode Elapsed time must be recorded by the operator using a stop watch or other time recording device Connection to the analog output using a c...

Page 16: ...ohol Use clean cotton or latex gloves when directly handling these spacers 3 Position the ionizing air gun above the charge plate and activate the air gun blowing ionized air at the plate The timer in...

Page 17: ...tatic decay time measurement and storing of the test data is not required depress the B button The display will briefly indicate BREAK and then return to the DECAY TIME START menu Demonstration 7 Char...

Page 18: ...amaged Do not attempt to modify or repair no user serviceable parts inside contact 3M Service for repair To reduce the risks associated with fire or explosion Do not operate the analyzer in an explosi...

Page 19: ...probe setup zeroing Use 1 Remove the red cover from the remote field sensor probe 2 Position the sensor in front of the charged surface or object to be measured Important Note When measuring an unkno...

Page 20: ...der requirements Internal resistance Rin greater than 1000 ohms Connect recorder to the red and black banana jacks at the rear of the analyzer using banana plug style wires PC Requirements for Data Re...

Page 21: ...e cup electrode a 4 Connect remaining two wires between the digital voltmeter voltage inputs and the red black banana jacks analog output at the rear of the analyzer Select meter scale 2vdc 5 Select t...

Page 22: ...____________ Range Setting P S Voltage Display Voltage Pass Fail Analog Output Pass Fail 25 VDC 25Vdc Full Range 2 5 24 4 25 6 Full Range 2 0Vdc 10 100VDC 100Vdc Full Range 2 5 97 5 102 5 Full Range 2...

Page 23: ...field sensors should be covered at all times with the supplied red covers Extreme care should be taken to prevent contact to the sensor elements They should be kept free of insulating coatings caused...

Page 24: ...rized of cer of 3M Warranty Limited Remedy Limited Liability This product will be free from defects in material and manufacture for a period of one 1 year from the time of purchase 3M MAKES NO OTHER W...

Reviews: