background image

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

S

S

W

W

-

-

D

D

 

 

 

Weighing Indicator

 

 

F

F

o

o

r

r

 

 

S

S

u

u

p

p

e

e

r

r

 

 

W

W

a

a

s

s

h

h

d

d

o

o

w

w

n

n

 

 

S

S

c

c

a

a

l

l

e

e

s

s

 

 

 

   

1WMPD4002414 

Summary of Contents for SW-D

Page 1: ...S SW W D D Weighing Indicator F Fo or r S Su up pe er r W Wa as sh hd do ow wn n S Sc ca al le es s 1WMPD4002414 ...

Page 2: ...injury 2011 A D Company Limited All rights reserved No part of this publication may be reproduced transmitted transcribed or translated into any language in any form by any means without the written permission of A D Company Limited The contents of this manual and the specifications of the instrument covered by this manual are subject to change for improvement without notice ...

Page 3: ...7 CALIBRATION 10 7 1 Items of the calibration mode 10 7 2 Setting of the weighing capacity division decimal point position 12 7 3 Gravity acceleration correction 13 7 4 Calibration using a weight 14 8 SETTING FUNCTIONS OR UNITS FOR USE 16 8 1 Setting of the CF function 16 8 2 Setting unit for use 17 8 3 CF Function list 18 9 OPTIONS 19 9 1 SW 11 Display stand Wall mount 19 10 MAINTENANCE 20 10 1 N...

Page 4: ...area it might cause some interference and under these circumstances the user would be required to take at his own expense whatever measures are necessary to eliminate the interference FCC Federal Communications Commission in the U S A Classification of protection provided by enclosures The equipment is designed to comply with the IP Code of IEC 60529 The IP69K is explained as follows IP Internatio...

Page 5: ...ad cells is available Available range selection for the display resolution is 1 20 to 1 60 000 Free setting of the weighing capacity within the display resolution of 1 60 000 is available Available weighing units are kg kilogram g gram Ib pound oz ounce and pcs pieces for the counting mode The counting mode easily counts the number of objects of the same weight The comparator mode compares the dis...

Page 6: ...r all of the following items are included Model label 1083013259 2 SW D CAPACITY SECTION Write the scale s ratings etc SW series INSTRUCTION MANUAL SW series Instruction manual SW D SET UP INSTRUCTIONS SW D Set up instructions this manual Main power cord Please confirm that the main power type is correct for your local voltage and receptacle Indicator To be attached ...

Page 7: ...d cables for all connections Ground the platform to be connected to the indicator to avoid a risk of electric shock Read the section PRECAUTIONS in the instruction manual of the SW to understand the conditions and precautions for installation operation and cleaning the indicator 5 2 Connecting to the load cell 1 Disconnect the main power cord of the scale from the power source 2 On the bottom of t...

Page 8: ...ng a weighing scale the following calculation formula should be satisfied The input sensitivity requires a change in output voltage from the weighing device to change the display value as a one digit step In order to achieve a scale with stable performance select the input sensitivity as large as possible Calculating formula A Rated capacity of load cell B Rated output mV V d Minimum division E Ex...

Page 9: ...herefore there is no problem with this design 5 4 Setting Up the Indicator After confirming that nothing is touching the key switches on the indicator connect the main power cable to a power source that has an earth ground terminal 5 5 Parameter Settings and Calibration When a platform scale is initially set up it is necessary to set some parameters and calibrate it using a calibration weight See ...

Page 10: ...splayed Turns on when the comparator results are displayed Weighing units pcs g kg oz and lb are available A selected unit turns on LO OK HI pcs g kg oz lb STABLE NET ZERO SAMPLE ON OFF PRINT UNITS ZERO TARE SAMPLE key LED display Weighing units STABLE indicator ZERO indicator ON OFF key PRINT key TARE key NET indicator UNITS key ZERO key Earth terminal Main power cord Load cell connector Display ...

Page 11: ...ss and hold to go to the comparator value setting mode PRINT key Outputs the weight value to the printer In the setting mode this key is used to increase the value of the selected blinking digit by one UNITS key Switches the weighing unit In the setting mode this key is used to shift the blinking digit to the right ZERO key Zeroes the scale and sets the display to zero TARE key Subtracts the tare ...

Page 12: ...ion Gravity acceleration correction Calibration using a weight How to enter the calibration mode 1 Make sure that the scale is in the weighing mode displaying kg g lb oz or pcs 2 Remove the screw protection caps on the load cell panel on the bottom of the display Loosen the four screws and then open the load cell panel The calibration CAL switch is located inside 3 Press the CAL switch The indicat...

Page 13: ...xecute it To exit from the calibration mode press the ZERO1 key or the CAL switch The indicator turns the power off Or press the ON OFF key to turn the power off Weighing mode Gravity acceleration correction Calibration using a weight Setting of the CF function Power OFF Press or Press the CAL switch Setting of the weighing capacity minimum division and decimal point ...

Page 14: ... lb Press the TARE key to store the currently set minimum division and calibration unit and proceed to the next step 4 After displaying Cap the indicator displays the currently set weighing capacity and decimal point position Display example 0060 00 kg Set the weighing capacity and decimal point position using the following keys MODE Shift the blinking digit to the right PRINT Increases the number...

Page 15: ...1 Refer to 7 1 Items of the calibration mode to enter the calibration mode The indicator displays Cal5et 2 Press the UNITS key to display g 9 7985 gravity acceleration value 3 Press the TARE key to enter the gravity acceleration value setting mode 4 Change the displayed value using the following keys UNITS To shift the blinking digit to the right PRINT To increase the value of the blinking digit b...

Page 16: ... the calibration mode The indicator displays Cal5et 3 Press the UNITS key twice and display Cal 4 Press the TARE key Cal 0 is displayed Confirm that nothing is placed on the weighing pan and wait for the STABLE indicator to turn on 5 Press the TARE key The scale calibrates the zero point and displays the value of the calibration weight SPAN calibration The calibration weight value is equal to the ...

Page 17: ...layed After displaying Cal the setting is complete 9 To finish the setting procedure press the ZERO1 key or the CAL switch The indicator turns the power off Note If the scale will be moved to another location set the gravity acceleration value for the new location and calibrate the scale according to the procedure above Refer to the previous section to set the gravity acceleration value Press ...

Page 18: ...he following keys UNITS Proceed to next setting item PRINT Increases the number 1 at the blinking digit Change the parameter When the currently stored set is changed the STABLE indicator turns off Changing the parameter using the PRINT key at CF1 2 8 or 12 is not available 5 Press the TARE key to store the setting After displaying end the indicator displays Unit and proceeds to the unit setting mo...

Page 19: ...ch of the display digits are assigned to a unit setting The setting at the first and fourth digit is not available Display assignment 7 6 5 4 3 2 1 Unit pcs g kg Set disabled oz lb Set disabled The unit during selection is indicated With each digit if setting to 0 its unit is not available and if setting to 1 its unit is available Only when set to 1 is a unit available Even if a unit is stored ava...

Page 20: ...Tare is not cleared by zero operation Tare clear by zero operation Cf 7 1 Tare is cleared by zero operation Used internally Cf 8 3 Not available for setting change 0 Non limit Minus and printing limit Cf 9 1 Limit has function Refer to below note 0 No auto tare function Auto tare function Cf10 1 Auto tare function When Cf10 0 at at t at f has no function 0 Not available for setting of function 5t ...

Page 21: ...using the four screws provided 2 Adjust the display angle using the two clamps on the SW 11 After loosening the two clamps on the SW 11 adjust the display angle and re tighten the two clamps 3 If you want to use the SW 11 as wall mount bracket after fixing only the stand part of the SW 11 on the flat wall by using the screws attach the display to the SW 11 These screws are not provided For details...

Page 22: ...e error When setting Cf 9 to 1 indicates that total weight is 19d or less d minimum division In the case of turning on the stable indicator setting the display to zero by pressing the ZERO key is available Unit weight error Indicates that the sample weight is too light to set the unit weight in the counting mode CAL error Indicates that the calibration procedure is canceled because the calibration...

Page 23: ...ow red LED Display update Approximately 10 times per second Operating temp 10 C 40 C 14 F 104 F Power supply AC main 100V 10 15 50 60 Hz 20VA or SLA Battery option available use time approximately 70 hours Main power cord length Approximately 3 m Dimension 255 W x 119 D x 179 H mm Weight Approximately 2 1 kg In the case of one 350 ohm load cell no other option and turning on comparator LED The ava...

Page 24: ...22 11 2 External dimensions When attaching to SW 11 Unit mm 255 20 5 158 118 9 50 52 5 160 52 5 47 5 47 5 4 M4 depth 8mm 315 283 152 5 197 5 250 90 70 6 5 φ 135 9 110 160 50 65 4 M4 x 8 truss screw ...

Page 25: ... m s2 Cape Town 9 796 m s2 Rio de Janeiro 9 788 m s2 Chicago 9 803 m s2 Rome 9 803 m s2 Copenhagen 9 815 m s2 San Francisco 9 800 m s2 Cyprus 9 797 m s2 Singapore 9 781 m s2 Djakarta 9 781 m s2 Stockholm 9 818 m s2 Frankfurt 9 810 m s2 Sydney 9 797 m s2 Glasgow 9 816 m s2 Taichung 9 789 m s2 Havana 9 788 m s2 Tainan 9 788 m s2 Helsinki 9 819 m s2 Taipei 9 790 m s2 Kuwait 9 793 m s2 Tokyo 9 798 m s...

Page 26: ...24 World map ...

Reviews: