CALIBRATION INSTRUCTIONS (Ver. 5, 6, 7 Software) 

Note: 

Below, position

 Weight A kg

 refers to the switch position furthest away 

from the test/data socket. 

Weight A lb 

refers to the switch position next to 

Weight A kg

 

1.

 

NO calibration Jack is required for this version 5,6 and 7 UC321. 

2.

 

Set the mode switch on the bottom of the UC321 to the position furthest 
away from the test/data socket. 

3.

 

Press the 

Blue Measure Switch

 once then slide the mode switch towards 

the test/data socket, (all the way) and then back again

Within two seconds. 

4.

 

If the display shows “

- - - - -

”, press 

Measure Switch

 three times. 

5.

 

After the display turns off then back on the display will show “

- - - - -

” again. 

6.

 

Press 

Measure Switch 

momentarily. The display will show “

99999

99999

99999

99999

” and start 

to count down. Press 

Measure Switch

. The display will show “

0000

 

7.

 

To check the calibration type set the mode switch to the position furthest 
away from the test/data socket and press the measure switch. 
By default, scale is set to 

llll----1111

, which uses 3 point calibration. 

To change set mode switch to the next position towards the test/data socket 
then press 

Measure Switch 

momentarily. Use

 Measure Switch

 to change 

setting. 

llll----0000

:2 point, (0kg,150kg). 

llll----1111

:3 point, (0kg,80kg,150kg). * DEFAULT. 

llll----2222

:2 point, (0lb,350lb). 

llll----3333

:3 point, (0lb,176.4lb(80kg),350lb). 

Slide Mode switch to position furthest away from the test/data socket press 

Measure Switch 

repeatedly to show 

5555

 

8.

 

Press 

Measure Switch 

momentarily, the display will show the temperature 

at calibration; i.e. (

20

20

20

20

). This must be set before calibration. 

9.

 

To change set mode switch to 

Weight A lb

 position then press 

Measure 

Switch, 

press 

Measure Switch 

momentarily, the temperature value will 

increase one digit at a time.

 

When the correct temperature is displayed

 

slide 

Mode switch to 

Weight A kg

 press 

Measure Switch 

display “

6666

” 

10.

 

Press 

Measure Switch 

momentarily, the display will show the gravity 

setting applied at time of calibration, i.e. (

9798

9798

9798

9798

). 

11.

 

To change set mode switch to 

Weight A lb

 position then press 

Measure 

Switch,

 

12.

 

Press 

Measure Switch 

momentarily to increment the value one digit at a 

time. Once the correct value is displayed, slide Mode switch to 

Weight A kg

 

press 

Measure Switch 

press 

Measure Switch 

to display “

7777

” 

13.

 

Press 

Measure Switch 

momentarily, the display will show the gravity 

setting for the location of use of the scale, i.e. (

9798

9798

9798

9798

). If the scale is to be 

used at the same location, where it is calibrated, 

this value must be set 

exactly the same as above!

 

14.

 

Use the same procedure as above to set the gravity value. 

 

……..cont. page 2