BATTERY REPLACEMENT PROCEDURE 

 

POWER & ENERGY LOGGERS 

MODELS PEL 102 & PEL 103 

(Battery for Cat. #’s 2137.51, 2137.52, 2137.61, 2137.62) 

 

The Power & Energy Loggers PEL 102/103 include an 8.4V NiHM 
battery pack 

(Cat. #2137.81)

 for backup power. This battery can 

be replaced by the user. The battery should only need to be 
replaced approximately every three or four years. This battery is 
available for purchase directly on our storefront at: 

www.aemc.com/store 

 

 
 
 
 
 
 
 

Summary of Contents for PEL 102

Page 1: ...Y LOGGERS MODELS PEL 102 PEL 103 Battery for Cat s 2137 51 2137 52 2137 61 2137 62 The Power Energy Loggers PEL 102 103 include an 8 4V NiHM battery pack Cat 2137 81 for backup power This battery can be replaced by the user The battery should only need to be replaced approximately every three or four years This battery is available for purchase directly on our storefront at www aemc com store ...

Page 2: ...ing A clean work surface free of debris water and other possible contaminants This surface should be well lighted and provide sufficient room to perform the steps below 8 4V NiHM battery pack Torx T10 screwdriver Torx type screwdrivers are also known as star screwdrivers Tape masking electrical etc Optional A small non conducting implement such as a drink stirring stick or toothpick Before startin...

Page 3: ...r probes Then follow these steps 1 Apply tape to the control buttons on the PEL front panel see Figure 1 This tape will ensure that the buttons remain in place when you remove the PEL s front cover from the back cover 2 Tape the PEL s endplates to the back cover see Figure 2 This helps to ensure that the endplates remain attached to the back cover when you remove the front cover Figure 1 Figure 2 ...

Page 4: ...te the six fastening screws on the back of the instrument Figure 3 4 Use a Torx T10 screwdriver to remove these fastening screws 5 Carefully separate the front cover of the PEL from the back cover Typically the two covers fit together tightly so it may be difficult to separate them at first ...

Page 5: ...urn starting with the bottom right corner With your fingers on the bottom endplate pull the covers apart while ensuring that the endplate remains attached to the back cover CAUTION Do not allow either the top or bottom endplate to separate from the back cover This may cause damage to the input and connection terminals that would require the PEL to be sent to AEMC for repair Figure 4 ...

Page 6: ...Model PEL 102 103 Battery Replacement Procedure Cat 2137 81 6 6 With the front cover removed locate the battery pack Figure 5 ...

Page 7: ...Gently remove and unplug the battery from the PEL A small non conductive implement such as a drink stirring stick shown below can be useful for this step Figure 6 8 Inspect the plug of the replacement battery back and note the position of the connector pins Figure 7 ...

Page 8: ...nsert the plug by gently pushing it into the socket within the PEL There is very little room for this procedure so a non conducting implement such as a stirring stick or toothpick may be helpful in ensuring the plug is securely inserted WARNING The PEL instrument will likely turn on when you plug in the connector Please ensure that you do not touch exposed circuitry as it is now energized and can ...

Page 9: ...it within the PEL When doing this ensure that the battery s connector wires are positioned as shown in Figure 9 below Be sure that the wires do not cross over the top of the post as shown in Figure 10 below If the wires are placed over the post the wires will likely be damaged when you replace the PEL top cover Figure 9 Figure 10 ...

Page 10: ... cover while carefully ensuring that it is securely in place 12 Turn the PEL over and replace the six fastening screws removed in step 3 of this procedure 13 Remove the tape applied to the front cover and endplates at the beginning of this procedure The PEL is now ready for operation ...

Page 11: ...NOTES ...

Page 12: ...11 21 99 MAN 100410 v4 Chauvin Arnoux Inc d b a AEMC Instruments 15 Faraday Drive Dover NH 03820 USA Phone 603 749 6434 800 945 2362 Fax 603 742 2346 www aemc com ...