Alcohol Analyzer 

Operating Manual 

V1.0 

 

 

AlcoEasy A65

 

Summary of Contents for AlcoEasy A65

Page 1: ...Alcohol Analyzer Operating Manual V1 0 AlcoEasy A65...

Page 2: ...6 5 5 1 Requirements 6 5 5 2 Measuring 6 5 5 3 Printing 7 5 6 Switch off 8 5 7 Error message 8 5 7 1 Blow failure 8 5 7 2 Low battery 9 6 DATA STORE 9 7 MAIN MENU 9 7 1 Active mode 10 7 2 Passive mode...

Page 3: ...Shenzhen Intermed Co Ltd can be used for repairs The sensor used in the device contains chemical material and must be disposed according to local regulations 3 PRODUCT AND FUNCTION The AlcoEasy A65 is...

Page 4: ...ont view 4 2 Icon Symbols Description Symbols Description Capacity indicate 29 5 Environment temperature 17 31 Clock Back Up button Main menu Down button Power switch NO Description 1 Mouthpiece port...

Page 5: ...should be clear that the device was calibrated before use If it was expired for calibration the device can t be used 5 3 Operating on active mode With active mode the device can measure the alcohol w...

Page 6: ...ng the measurement It should be clear that the person to be tested has not drunk any alcohol just before measuring is taken Otherwise the test should wait at least 15 minutes The mouth freshener syrup...

Page 7: ...ow as below 17 31 01 08 Active mode 1 TESTING The time of automatic process is decided to the concentration of the breath alcohol 5 5 3 Printing After the air has been analysed the reading is apper th...

Page 8: ...8 Active mode 1 BAC No Date Time CN PN LN 00011 2016 01 08 17 31 Input Pr int 5 6 Switch off By pressed down the power button for approximately 3 sec The divice will have following display 17 31 01 08...

Page 9: ...n the flash memory will not be lost even there is no power on the device The flash memory will be overwritten if it is full but there will have an alarm before overwritten The first data will be overw...

Page 10: ...ctive mode you will hear beep sound during blow and it will not stop untill blowing volume reached 1 0 L The screen will show Main menu 17 31 01 08 OK Settings 4 Results 3 Active mode 1 Passive mode 2...

Page 11: ...o Date Time CN PN LN 00011 2016 01 08 17 31 Input Pr int If active mode detect alcohol C 80mg 100ml it will show as follow and indicator light will show red mg 100mL 95 17 31 01 08 Active mode 1 BAC N...

Page 12: ...ive mode 2 17 31 01 08 Sampling 2 Passive mode Start Notice No need to use mouthpieces during passive mode Put the device close enough to the nose or mouth to the tester press Start button the device...

Page 13: ...Press button it will show the next record and you can print previous test record freely 7 4 Setting On the Settings menu you can change or define some functions of the device After select setting menu...

Page 14: ...number plate the numbers of the date and time can be modified 7 4 2 Units Select Unit menu press ok button it will display the optional units BAC means blood alcohol concertration BrAC means breath a...

Page 15: ...umber 5 Function 4 3 Cali Remind Air blank set 1 2 Print setting 3 Cali Remind 17 31 01 08 Result mode 4 OK Print Number 5 Function 4 3 Cali Remind Air blank set 1 2 Print setting 3 17 31 01 08 OK 1 C...

Page 16: ...2 OK 5 Function 4 3 4 3 Print Number Result mode Print setting 17 31 01 08 OK 4 Print Number Number 1 Result mode 17 31 01 08 Cali Remind Air blank set 1 2 OK Function 4 3 4 5 3 Print setting Result m...

Page 17: ...ssword only authorized person can reach this menu 17 31 01 08 Date Time Units Function TP Cali 4 1 4 2 4 3 4 4 OK Settings 4 Services 4 5 17 31 01 08 1 2 3 4 5 6 7 8 9 0 Enter PIN Services 4 5 OK 8 Cl...

Page 18: ...Calibration should operate by authorized person 10 Battery Charge 10 1 Capacity Indicator AlcoEasy A65 battery rate can know clearly via the battery Indicator on the left top screen If battery too low...

Page 19: ...Over 11 Error and Remedy Error message Causes Remedy 17 31 01 08 Active mode 1 Blow Failure Blow was interrupted Blow again and reach to 1 0L Time Error OK Date and time error If this error happens a...

Page 20: ...ase shake device Make another measure if this error appear again get help from service people Memory Error OK Memory can not be read or write It s possible that new date can not be stored and the old...

Page 21: ...LxWxH Weight Active and Passive Mode 0 550 mg 100ml BAC 4 4 mg 100ml 5 absolute 20 absolute 0 2 5 mg L BrAC 0 020 mg L 5 absolute 20 absolute 5 C 50 C 30 C 70 C 0 C 25 C 600 1400 hPa 20 98 r h 1 0 L 3...

Page 22: ...n material and its quality problems in 1 year after purchase we will repair the device for free regarding warranty regulations If the device damaged or can not work for Man made causes operation negli...