2352 Walsh Ave. Santa Clara, CA 95051. U. S. A. Tel.: (408) 748-9100, Fax: (408) 748-9111

   

www.analogtechnologies.com 

 

Copyrights 2000 – 2012, Analog Technologies, Inc. All Rights Reserved. Updated on 11/22/2012   

1

 

 

Analog Technologies 

      VC60B

 

              Multimeter 

 

Figure 1. The Photo of Actual VC60B 

 

    Figure 2. The Photo of Actual VC60B 

 

Figure 3. The Photo of Actual VC60B 

 

FEATURES 

Altitude: <2000m   

Power Consumption: <650mW 

Sampling Rate: Approx. 3 times/sec. 

Fuse: 100mA/60V Resettable Fuse 

Lower Battery Indication: The LCD Display “

” Symbol. 

Over Range Indication: Displays “1” When Exceeding 
Upper Limit 

Additional Temperature Coefficiency: 0.15 × Specified 
Accuracy/

 (<18

 or >28

Power Source: 5# Battery LR6 (1.5V) × 6 (or External AC 
Adapter). 

APPLICATIONS

 

It’s widely used for insulation testing of electric machines, 
cables, electronic equipments, electric facilities, etc. 

DESCRIPTION 

 

VC60B+1.0 is a digital insulation tester which adopts DC 
voltage converter of low consumption and high converter 
ratio inductance energy storage to convert the voltage of 
9Vdc into 250V/500V/1000Vdc.   

With digital bridge, the instrument has the features of being 
easy to operate, wide test range, highly stable performance, 
backlit LCD display, data hold and auto power off. With a 
shoulder strap, it can be used by both hands.   

It is an ideal instrument to be used for insulation testing of 
electric machines, cables, electronic equipments, electric 
facilities, etc. 

OPERATION INSTRUCTION

   

For the first time of operation, please make sure there is 
enough power source. If not, open the battery door and put 
into 6 pcs of 5# batteries. Please pay attention to the batter 
polarities. 

When connecting to an external AC adapter, the battery 
power supply will be shut off. Do not charge the batteries. 

Note:

 choose a correct power supply mode (

). 

 

 

 

 

 

 

 

Summary of Contents for VC60B

Page 1: ...tronic equipments electric facilities etc DESCRIPTION VC60B 1 0 is a digital insulation tester which adopts DC voltage converter of low consumption and high converter ratio inductance energy storage to convert the voltage of 9Vdc into 250V 500V 1000Vdc With digital bridge the instrument has the features of being easy to operate wide test range highly stable performance backlit LCD display data hol...

Page 2: ...r 28 Operation Environment Temperature 15 55 Relative humidity 75 RH Storage Environment Temperature 40 60 Relative humidity 90 RH LCD Size 84 8 59 8mm Max Display 1999 Dimension 175 L 110 W 62 5 D mm Weight 580g including battery TECHNICAL SPECIFICATIONS Table 2 Basic Function Range Basic Accuracy Output Voltage 250V 500V 1000V 10 Testing Current 250V R 250kΩ 1mA 500V R 500kΩ 1mA 1000V R 1MΩ 1mA ...

Page 3: ... terminal for tested circuit 4 L Terminal for connection to the GND of the tested object 5 G Protection terminal When the tested object is required to add the protection loop to eliminate the leakage effect connect the electrode wire of the protection loop to G terminal 6 Voltage measurement resistance measurement 2kΩ and continuity measurement input terminal V Ω 7 Test button 8 High voltage indic...

Page 4: ...the ground terminal of the tested object Lead E is required to be hung in the air 6 If the LCD displays 1 it indicates over range operation Please choose a higher range to get reading Caution 1 Pressing the TEST button output terminal E will have high voltage output Be careful to avoid electric shock 2 Before testing make sure to choose a correct voltage range Make sure the LCD display is in accor...

Page 5: ... the instrument circuit 3 During high voltage circuit measurement do not touch the high voltage circuit to avoid electric shock DATA HOLD Press the HOLD button the current data will be held on the LCD Press the HOLD button again to cancel data hold Note There is no data hold function during insulation resistance measurement BACKLIT DISPLAY Pressing button the LCD backlight turns on Pressing button...

Page 6: ...d replace with a new one To ensure long time battery power supply it is recommended to use alkaline batteries 3 Load the battery door and tighten the screws PACKING LIST 1 Digital insulation tester x1pcs 2 10A test lead x1 pair 3 5 battery 1 5V x6pcs 4 Operation manual x1pcs 5 Shoulder strap x1pcs 6 Crocodile clip x1 pair TROUBLE SHOOTING If the instrument does not work properly the following self...

Page 7: ...e is not necessarily performed except those mandated by government requirements 3 Customers are responsible for their applications using ATI components In order to minimize risks associated with the customers applications adequate design and operating safeguards must be provided by the customers to minimize inherent or procedural hazards ATI assumes no liability for applications assistance or cust...