Instructions for use

1   Product

1.1 Product description

Table of contents

1  Product

2  Certificates / Declaration of conformity

3  Labelling

4  Preparation of commissioning and use of product

5  Disassembly of used filter / filter change

6  Accessories

7  Maintenance / cleaning

8  Storage

9  Return

7 Maintenance and cleaning

The Germlyser

®

 W requires no maintenance within the specified operating time. The outside of the filter 

can be cleaned using common surface disinfectants.

When replacing the Germlyser

®

 W, a small amount of unfiltered water will flow out. Over time, this will lead 

to deposits on and in the adapters. Aqua 

free

 recommends replacement of the adapters after a period of  

6 to 12 months.

In the case of a deviation from the specifications or a damage to the packaging, the product should be 

returned to Aqua 

free

, along with a Defect Report (form available to download on the service page of the 

Aqua 

free

 website www.aqua-free.com).

Please note:

 Aqua 

free

 operates its own replacement service, which can be used by facilities with 25 or 

more filterpoints of use.

8 Storage

The Germlyser

®

 filter must be properly stored in a dry and hygienic environment at a temperature between 

10 and 50°C.

9 Return

The Germlyser

®

 reusable filters must be sent back to Aqua 

free

 within the agreed time for reprocessing 

(form downloads at the service page under www.aqua-free.com).

GEBR-ANW-W-GB-0

5

/20

21

-V00

3

Germlyser

®

 W

The Germlyser® W is a      -marked, point-of-use sterile shower filter for use in birthing pools and burn baths 

to  prevent  bacterial  nosocomial  infectious  diseases  caused  by 

Pseudomonas  aeruginosa,  Legionella 

pneumophila,  Stenotrophomonas  maltophilia 

or

  Staphylococcus  aureus

  among  immunosuppressed  and 

immunocompromised patients.

     

1.2 Technical data

Size: 

L 206 mm x Ø 43 mm

Retention: 

7 log steps 

Brev. dim.

Pore size: 

0.2 µm

Operating time*: 

4 weeks

Chlorine resistance**: ≤10 ppm 

Flow rate: 

20 l/min at 5 bar, at 25°C 

Max. operating pressure***:  5 bar at 60°C 

Max. operating temperature:  60°C (at 70°C < 30 min. 

over operating time)

Connection: 

Quick-release coupling

Headquarters:

Aqua free GmbH

Winsbergring 31
22525 Hamburg
Germany

Phone:

  +49 (0)40-46 89 99 90

Fax:

    

+49 (0)40-46 89 99 99

info@aqua-free.com

www.aqua-free.com

Aqua free Solutions Ltd.

The Media Centre
7 Northumberland Street

Huddersfield, HD1 1RL, Great Britain

E-Mail:

info@aqua-free.co.uk 

Phone:

+44 (0)1484 483045

*  The operating time can be increased up to 8 weeks if sufficient hygienic measures exist and an adequate 

hygiene plan is carried out. We recommend monitoring by appropriate clinical microbiological testing.

**  Continuous dosage of ≤ 10 ppm over operating time; short-term (1h) high dosage (400,000 ppm) 

 

for chemical disinfection.

*** The Germlyser

®

 W is not suitable for connection to low-pressure systems (e.g. boilers).

Reviews: