Initial Equipment Check
• Make sure the Ranger unit’s
power cord is plugged into the unit.
• When checking for grounding, clamp to the
screw threads on the unit’s IV pole clamp or the
equipotential stud on the rear of unit.
Clamp should be tightened securely.
• Sometime during shipping, this unit may
have been exposed to extreme temperatures.
The unit may need a period of 20 minutes or
more to return to normal operating temperature.
Store the Ranger unit at room temperature,
allowing it to equilibrate, before using
or testing the unit.
*
Check the Ranger system website to ensure you have the most recent version of this document. www.rangerfluidwarming.com