INSTRUCTION MANUAL

Technical changes reserved 

Description

B+B 0141 0315-117 September 2010

B+B Thermo-Technik GmbH | Heinrich-Hertz-Straße 4 | D-78166 Donaueschingen 

Fon +49 771 83160 | Fax +49 771 831650 | info@bb-sensors.com | bb-sensors.com

01 / 12

Digital Quick-Response Thermometer

HM 1150

Summary of Contents for HM 1150

Page 1: ...served Description B B 0141 0315 117 September 2010 B B Thermo Technik GmbH Heinrich Hertz Straße 4 D 78166 Donaueschingen Fon 49 771 83160 Fax 49 771 831650 info bb sensors com bb sensors com 01 12 Digital Quick Response Thermometer HM 1150 ...

Page 2: ...stomer shall observe the following Operating Instructions In the case that against one s expectations any trouble occurs which you can not resolve yourself please contact our service centers or our authorized dealer We will provide you rapid and competent help to minimize the risk of long time outfalls The following operating Instruction is an indispensable part of this Product It contains importa...

Page 3: ...9 771 831650 info bb sensors com bb sensors com 03 12 Content Foreword 02 General Information 02 Symbols Employed 01 Warning Signs 04 Safety Instructions 05 Intendend Use 06 Disposal 06 1 Points to be observed during operation 07 2 Recalibration 08 3 Correctiontable typ value 09 4 Specification 10 5 Scope of Delivery 11 6 Optional Order Information 11 General Questions 12 ...

Page 4: ...sary for the success of a procedure They have to be followed in order to get good results Warning Signs Meaning This symbol advises the user of danger for persons material or environment The text gives information that must be necessarily followed to avoid any risks Caution against hot surfaces BGV A8 GUV V A8 W26 and hot liquids or substances Caution against liquids and hot substances Caution aga...

Page 5: ...sted to the ambient temperature before trying a new start up If device is to be connected to other devices the circuitry has to be designed most carefully Internal connection in third party devices e g connection GND and earth may result in not permissible voltages impairing or destroying the device or another device connected Warning Just devices with mains input If device is operated with a defe...

Page 6: ...reasons This instruction manual does not at all substitute any additional instruction manual of connected accessoriy This unit has been marked in accordance with the European Device 2002 96 EC on waste electrical and electronic equipment WEEE At the end of its useful operating life dispose of the unit as electrical scrap Please ask either B B Thermo Technik GmbH or your specialist dealer for infor...

Page 7: ...vice are always subjected to the same temperature i e try to avoid holding sensor plug in your hand for a longer periode of time as well as subjecting device to an additional heat source as this may result in measuring inaccuracies d Mains operation When using a power supply unit please note that operating voltage has to be 10 to 12 V DC Do not apply overvoltage Cheap 12V power supply units often ...

Page 8: ...rature possible to calibrate the device Calibration point 0 C Put ice cubes in a glass and pour cold water till ice cubes are almost covered Put sensor into glass wait approx 15 minutes then stirr water with a spoonhandle Wait for stable value to be displayed then turn zero point potentiometer NP Potentiometer next to sensor connection by means of a screw driver till display shows 000 Calibration ...

Page 9: ...20 20 230 228 440 441 650 659 860 871 1070 1073 30 29 240 238 450 452 660 670 870 881 1080 1082 40 39 250 248 460 462 670 680 880 891 1090 1091 50 49 260 258 470 472 680 690 890 901 1100 1100 60 59 270 268 480 483 690 700 900 911 1110 1110 70 70 280 278 490 493 700 711 910 920 1120 1119 80 80 290 288 500 504 710 721 920 930 1130 1128 90 90 300 298 510 514 720 731 930 940 1140 1137 100 100 310 308 ...

Page 10: ...mperature changes if possible otherwise a temperature adjustment time of approx 15 minutes has to be taken into account 0 to 80 r F not condensing Power supply 9V battery type IEC 6F22 included as well as additional d c connector internal pin Ø 1 9mm for external 10 12V direct voltage supply suitable power supply GNG10 3000 Battery life time approx 700 operating hours Low battery warning BAT displ...

Page 11: ...er HM 1150 SET incl 1 x Type K Probe with 130 mm nominal length and 3 mm diameter in a transportation case Articlename Articelnumber Description Immersion Probe type K for extreme use 0600 1205 Diameter 6 mm nominal length 300 mm Stainless steel B B exklusive handhold cabel length 1000 mm with helix cable Miniatureplug yellow Measuringtip swagged diameter 3 0x15 mm Measuring range 50 850 C Penetra...

Page 12: ...h Hertz Straße 4 D 78166 Donaueschingen Germany Tel 49 0 771 83160 Fax 49 0 771 831650 E Mail info bb sensors com bb sensors com We wish you a successful measuring Your Temperature Partner B B Thermo Technik GmbH All technical information s in this document are proved by us with high accuracy and shall inform you about all application possibilities This information s are not confirmed by us and ne...