Series Evolution

Evolution Series

I n s t r u c t i o n s

Summary of Contents for Evolution Series

Page 1: ...Series Evolution Evolution Series I n s t r u c t i o n s...

Page 2: ...2...

Page 3: ...ction and the highest level of safety We thank you for the confidence you have shown us Our company guarantees you that we select materials with care and conduct quality controls of state of the art p...

Page 4: ...15 Image 16 One piece bed rail 16 Split bed rail 17 Function description 18 Backrest RL 18 Mechanical unlocking of the backrest option 18 Upper leg section OT 18 Height adjustment HV 18 Comfort sittin...

Page 5: ...ses 29 Side Rail Extender Three Side Rails 29 Troubleshooting 30 Reference test only for Standard ICS 31 Mains isolation 31 Troubleshooting 32 Product safety 33 Cleaning and Disinfection 34 Disinfecti...

Page 6: ...6 Evolution 100 Evolution 200 Evolution 300 XL Evolution 400 Evolution 400 LR100 P R O D U C T D E S I G N A T I O N...

Page 7: ...for use in old age and nursing care environments These beds are used exclusively for this purpose The beds convenience and func tionality provide relief for personnel when performing nursing duties an...

Page 8: ...lowed The instruction booklet is designed for the Evolution series of patient beds The illustrations diagrams and texts included do not cover the entire line For simple orientation in these instructio...

Page 9: ...ts 9 The function of the side rails must be inspected daily and they may not be loaded more than 70 kg in vertical po sition or 50 kg in horizontal position 10 When lowering the mechanical emergency b...

Page 10: ...verheated operations are automatically swit ched off The motor can be switched on again only after a 2 hour cooling down period Blocking the mechanical parts of the bed must be avoided as this can lea...

Page 11: ...ator directive must be included in an inventory list for each business premises It is re commended to document the proper execution of the required technical safety checks in the inven tory and to pre...

Page 12: ...kg 170 kg patient weight capacity 30 kg mattress 25 kg accessories Heavy duty bed 300 XL 300 kg 250 kg patient weight Weight approx 120 kg Heavy duty bed 300 XL approx 160 kg Protection type IPX4 Lyi...

Page 13: ...technical testing after delivery ac cording to VDE0751 1 Sample images of a care bed Evolution 400 nameplate in the standard design with mains connection 8 Declaration of the used symbols on the name...

Page 14: ...contour up S7 Lift adjustment up S9 Anti Trendelenburg S11 Lift position S13 Memory S4 Leg rest adjustment down S6 Auto contour down R1 LOCKED No functions operational R2 FREE All functions available...

Page 15: ...NT POSITION All functions available except S10 R4 FREE All functions available S12 Low positioning S1 Back rest adjustment up S2 Back rest adjustment down S3 Leg rest adjustment up S5 Auto contour up...

Page 16: ...g rest Sitting area fixed CPR emergency lowering optional Latch for bed rail Erectable bed rails one piece Frame bracket for trapeze bar Foot rest Chassis Evolution 100 Chassis Evolution 200 Foot peda...

Page 17: ...itting area fixed Bed rails telescopic CPR emergency lowering optional Erectable bed rails split Frame bracket for trapeze bar Foot rest Chassis Evolution 300 Chassis Evolution 400 Adjustable lower le...

Page 18: ...ight adjustment F U N C T I O N D E S C R I P T I O N Depending on the model and type of hospital bed the bed can be used in the following positions by adjusting the backrest the upper and lower leg s...

Page 19: ...itting positi on ensure the safety of the patient No foreign bodies must be in the movement area of the lifting mechanism Auto contour The adjustment for auto contour position is made using the respec...

Page 20: ...hand control unit and can be operated by personnel using a key switch In the event of a malfunction of the electric drive the functions can be stopped using the appropriate rotary switch Operation The...

Page 21: ...tions and under supervision of nursing personnel Ensure the safety of the occupant Evolution 300 The Evolution 300 model can be trans ported in any position is equipped with a central roller fixing wh...

Page 22: ...ution 400 LR100 offers 3 different adjust ment possibilities 1 Central caster locking pedal down 2 4 casters 360 free pedal in middle position 3 1 caster locked in directional position pedal up Cautio...

Page 23: ...ing and raising the side railings take great care not to jam fingers hands or other body parts between the side railings and the frame of the lying surface In the case of disoriented or undernourished...

Page 24: ...mobility support for a patient s movements 3 The side railing is unlocked for height adjustment using two pressure pieces in the inner upper area and positi oned to the maximum height Perform telesco...

Page 25: ...25...

Page 26: ...During use for example for purposes of exchange simple assembly procedures may also be performed by technically qualified personnel For professionally authorised personnel we of fer extensive service...

Page 27: ...as the back support in the intended mounts on the mattress frame The infusion stand is intended only for hanging infusions and not for attaching other accessories or the like Maximum load is 8 kg 2 k...

Page 28: ...one piece bed sides for safety reasons it is imperative that the existing safety elements and accessories are exchanged Urine Bottle Holder Urine Bottle Urine Bag Holder The mount for the urine bottl...

Page 29: ...ufacturer about the mattresses you need Important Use only mattresses approved and tested by Bakare Beds Ltd For reasons of safety the thickness of the mat tress and the standard safety spacing of the...

Page 30: ...er or directly our customer service on 01752 512227 Power Supply available and cables all pluged correctly LED at hand control unit X1 is red blinking Carrying out a reference test Remove foreign bodi...

Page 31: ...d flashing to green The system is now fully initialised Mains isolation For the safety of the resident Malsch care design offers mains isolation with their system Standard ICS This only supplies the d...

Page 32: ...ving parts and remove foreign bodies The safe load has been exceeded Reduce load Motor shuts off after continuous use Adjustment time or safe load has been exceeded motor reacted to excessive heat All...

Page 33: ...08 from 1 1 01 Adjustable beds for disabled persons As per DIN EN 60601 1 from 1 1 01 Medical electrical devices As per DIN EN 60601 2 52 2010 Medical electrical devices and accessories As per EN 1253...

Page 34: ...f Wood Parts For Bakare Beds Ltd beds all wood bodies are finished in compliance with the DIN 68861 1A standard The objective is to achieve a stylish design best functionality and practical use To ens...

Page 35: ...of the product only if it is regularly serviced and used in accordance with the safety guidelines If significant defects occur during maintenance so that safe operation of the product cannot be ensure...

Page 36: ...ree replacement or remedy within the warranty period Malfunctions or defects caused by improper handling or external factors are excluded from the warranty Our terms of business and delivery apply Our...

Page 37: ...for handicapped people DIN EN 60601 2 52 2010 reliability of electr Operating beds All products are developed produced and tested under applying the TQM system DIN EN ISO 9001 certificate no 73 100 12...

Page 38: ...teller Betten Malsch GmbH Rohbergstra e 9 D 36208 Wildeck Obersuhl Tel 49 0 6626 915 100 Fax 49 0 6626 915 116 dass das Produkt Pflegebett Serie Impulse den grundlegenden Anforderungen und Bestimmunge...