background image

                                                                                                                                                             

 

 

METALLOGRAPHIC MICROSCOPE

 

mod. MTM-1A 

 

 

USER MANUAL 

 

F i l e n a m e :   M T M - 1 A . d o c  

Summary of Contents for MTM-1A

Page 1: ...1 METALLOGRAPHIC MICROSCOPE mod MTM 1A USER MANUAL F i l e n a m e M T M 1 A d o c ...

Page 2: ...2 INSTALLATION OF MICROSCOPE Take out the field diaphragm adjustment screw from the package and plug each correspond of the hole Field diaphragm adjustment screw ...

Page 3: ...hing demonstration in the colleges and factory NOTE Stage cross travel range longitudinal 153 mm traverse153 mm for MTM 1 Stage cross travel range longitudinal 203 mm traverse 203 mm for MTM 1A Ⅰ SPECIFICATIONS 1 Eyepieces Type Magnification Focus mm Field mm Remark Wide field eyepiece 10X 25 φ18 16X 15 6 φ11 Optional Plan Dividing 10X 25 φ18 Optional 2 Objectives Type Magnification N A W D mm Rem...

Page 4: ... 220V 50HZ or AC 110V 60HZ 3 Turned filter group yellow blue green ground glass 4 With polarizer 5 Stage cross travel range longitudinal 153 mm traverse 153 mm for MTM 1 longitudinal 203 mm traverse 203 mm for MTM 1A 6 Fine focusing knob minimum division 0 0008mm 7 Interpupillary distance adjustment range from 53mmto 75mm 8 Anti fungus Yes ...

Page 5: ...aphragm adjustment screw 3 Polarization adjustment pole 4 Trinocular 5 Observation photo lever 6 Polarizer lever 7 Objective 8 Stage 9 Limit holding knob 10 Coarse focusing knob 11 Fine focusing knob 12 pole 1 2 3 4 5 6 7 8 9 10 11 12 Fig 1 ...

Page 6: ...er 5 Field diaphragm adjustment screw 6 Aperture diaphragm adjustment lever 7 Power supply plug 8 Electrical outlet 9 Light control 10 Power switch 11 Cross knob 12 Lengthwise knob 13 Adjustable tensional knob 14 Binocular locking screw Anti fungus installed 7 8 1 2 3 4 5 6 9 10 11 12 13 14 Fig 2 ...

Page 7: ...iaphragm lever 4 again to let the image of the diaphragm slightly larger than diaphragm in the field Fig 3 Now it s ready to use 4 This microscope is equipped with simple polarization device Push the Polarizer lever 3 to make the analyzer into the optical path of the observation Push the polarizer 2 into the vertical illuminator Fig 3 Rotate the polarizer 2 while watching the brightness in the fie...

Page 8: ...ol 1 to make the image clearly Fig 4 7 Coarse Fine focusing The instrument used coaxial coarse fine focusing mechanism The adjustable tensional knob 4 used for adjusting the tension of the coarse focusing knob 3 to prevent the stage from naturally sliding down The limit knob 1 prevents accidental specimen objective contact 2 is fine focusing knob Fig 5 8 Movement of stage Fig 5 The pole 5 could be...

Page 9: ...specimen clear H If brightness of field not equality you can adjust the screw 3 or move the lamp 2 slightly to make the brightness of field equality I Turn the field diaphragm to minimum adjust the field diaphragm adjusting screw to make the center of field diaphragm and field be superposition then turn the field diaphragm until you can not see the image of field diaphragm in the field Fig 2 J Whe...

Page 10: ...take out socket of fuse 1 Changing the fuse Install socket of fuse and plug again The specification of fuse Ф5 0 5A If stage dosen t balance you can adjust the adjustment nut of chassis Ⅳ MAINTENANCE 1 Sweep the lens Sweep the lens by lens tissue or soft fabric immersed with mixed liquid of alcohol ether or diethyl benzene Cleaning the oil on the 100X objective whenever finish operating 2 Clean th...

Page 11: ...embling the microscope Being a precise instrument do not disassemble the microscope casually that may cause serious damage to its performance 4 Being not used Cover the microscope with polymethyl methacrylate or polyethylene and places where there is dry and modules Suggest that storage all objectives and eyepieces in closed container with drying agent ...

Page 12: ...ng estop the field 1 Check the trinocular pole whether push to the locate point push to the locate point 2 Check the filter seat whether locate well put the filter seat good location D Can not see the image while adjusting coarse fine focusing See the instruction Ⅳ OBSERVING OPERATION 7 to adjust E Automatic offset the focus See the instruction Ⅳ OBSERVING OPERATION 7 to adjust F The image no clea...

Page 13: ...ive responsability that the Microscope models MTM 1A this declaration refers to are in compliance with the following rules EMC std EN 61326 1 2006 EN 61000 3 2 2006 EN 61000 3 3 1995 A1 2001 A2 2005 Harmonized by the directive 2004 108 EC SAFETY std EN 61010 1 2001 Harmonized by the directive 2006 95 EC ...