IR Microscope: Basic Instructions

 

 
Note: microscope manual available on BB. 
 
Log on to computer with university username  and password. 
 
Start OPUS and log on with password infrared. 
 
Fill IR detector Dewar with liquid nitrogen (normally use the one on the left 

– 

mid band detector). Fill it until it begins to spill out, and the red light under the 
“snowflake” symbol (K in diagram below) goes off. 
 

 

 
Checking the signal 
 
On the microscope, press the “eye” button (F in diagram). This gives a small 
spot (if y

ou don’t see it, you probably need to increase the intensity using the 

wheel below the stage - I in diagram). Focus it onto a mirror or, preferably, a 
less reflective bit of metal. Use the focus knob (on the right, at the back) to 
obtain a small spot. Now look through the binoculars and get a focused 
image, using the same knob.  
 
P

ress the “wave” button, on the left. 

 
In Opus, select Measure 

 Setup Measurement Parameters 

 
You can load a previous set of parameters by clicking Load (there is a file 
called Peter Basic Parameters in My Documents). 
 
Go to Check signal (tab at top right). 
 
You should get an 

interferogram somewhere 

around 70000. If you don’t see 

one, check you are in reflectance mode (under Advanced). Also make sure 
the 

detector is cold. 

 

When you have the interferogram, click 

Save Peak Position then Save and 

Exit. 
 