Quick Guide - V5 / V8 OXYGEN CONCENTRATOR USER GUIDE AND MAINTENANCE

 - Filter Change

 

 

 

V5 / V8 OXYGEN CONCENTRATOR REAR VIEW 

 

Air Inlet Filter access panel 

 

 

 
     

Exhaust                                                   Power cord  

(under base of unit) 

 

AIR-INTAKE FILTER ACCESS PANEL 
 

 

 

DUST FILTER  

 

 

 

 

 

 
 

AIR -INTAKE FILTER 

 

 
 

The Air-intake filter needs to be replaced after every 
1000 hours of use or annually (whichever comes first 
for the first change).  For subsequent filter changes the 
device will have a label indicating total working hours 
required before the next filter change.  The device 
needs returning to MESU for the filter changes.  
During home use a replacement air-intake filter is 
available from MESU. 

 

 
 

HUMIDIFIER BOTTLE 
 

Humidifier bottles are single patient use. 
Bottles must be cleaned daily with soap and water. 
 

    

    

 
 

Y TUBE CONNECTOR 

 

 
 
 
 

 

 
 

 
 

 
 

 

 
 
    

 

The dust filter is located on 
the rear of the Filter access 
panel door. 

 

 

 

 

 

 

The Dust filter can be cleaned 
with mild detergent and water, 
rinse to remove all residues and 
air dry. 

 

The dust filter should be 
inspected every 2 weeks and 
replaced when it is visibly dusty.  
 
 

 

The white air-intake filter is 
located in the air-intake 
filter compartment via the 
access panel. 

Dust Filter 

Air-intake 

filter 

Fill with distilled 
water only.   

 

Water must be kept 
between Maximum 
and minimum scale 
line 

 

Change the water 
every 24hrs 
 

  Administer 15lpm from 2 x V8s   

 

 

 