background image

PPM-1c / Part Per Million CO

2

 Monitor 

 

   The PPM-1c allows users to upgrade C.A.P.’s line of CO

2

 & Environmental controllers to control  

  CO

2

 with PPM / “Part Per Million” accuracy. The PPM-1c is capable of measuring CO

2

 PPM  

  levels from 0-5000. On-board adjustable set point and calibration makes it easy to use. 

   

  

• 

The PPM-1c is an extremely accurate Part-Per-Million (PPM) CO

2

 sensor. 

• 

It can measure CO

2

 from 0-5000 PPM +/- 100PPM. 

• 

It has an easy to read 4-digit LCD display.  

• 

It has an external power supply and a Quick Disconnect or “QD”, which is connected
to a CO

2

-2, CO

2

-4 Environmental controller or the CGC-1 Complete Greenhouse

Controller. 

• 

Once the PPM-1c is connected to either the CO

2

-2, CO

2

-4 or the CGC-1, the

controller will receive information from the PPM-1c and use it to control the CO

2

levels with extreme precision. 

• 

The PPM-1c works in harmony with the Inject and Sample timers on the CO

2

-2 ,

CO

2

-4 or the CGC-1. 

• 

The PPM-1c also has a visual Status indicator, which verifies the unit has completed
the “Initial warm-up” stage and is functioning properly. 

 
 
 
 
 
 
 
 
 

 
 
 
 

INSTALLATION 

 

1)  The base-plate must be mounted to a wall using the (4) screws provided with the unit. Find a location near a 120-volt 

receptacle for power. The PPM-1c will need to have a fair amount of airflow around the enclosure and be at “plant-level” 
to be most accurate.  

2)  Once the base-plate is mounted the sensor can be snapped onto the base-plate by hooking the two top tabs and then 

pivoting the unit down until it snaps into place.  

        *Note: To release the tabs, press in on the two snap tabs located on the bottom of the unit near each edge. 

3)  The PPM-1c comes pre-wired with a plug-in power supply. The power supply must be connected to a constant source of 

120-vac power. Once powered up, the PPM-1c will enter a “Initial warm-up” and calibration mode. Slowly over a 10-20 
minute period, the CO

2

 level should stabilize between 250-550 PPM outdoors and as much as 1500 PPM indoors. 

 
CONNECTING TO THE ENVIRONMENTAL CONTROLLER 

 

The other cable coming from the PPM-1c is the Quick disconnect or “QD”. The QD is a standard IEC cable which replaces 
the jumper plugged into the connector on the left side of the CO

2

-2 / CO

2

-4 and the bottom of the CGC-1. Once the PPM-1c 

is connected, the controller will automatically begin controlling CO

2

 with Part-Per-Million accuracy. 

 
MAKING ADJUSTMENTS 
 
 The PPM-1c has an adjustable setpoint from 0-5000 PPM. Once the desired setpoint is entered, the PPM-1c will disable the 
“CO

2

 Level Low” function of the environmental controller effectively shutting OFF the CO

2

 outlet. Adjusting the CO

2

 

setpoint and re-calibrating the unit is easy using the front mounted push buttons. There are four buttons. 

 

a) 

clear

: used with the mode button to enter the programming mode. 

b) 

mode: 

used with the clear button to enter the programming mode and to scroll through the different modes. 

c) 

up/down

: The arrow up / down button is used to change the program set points. 

d) 

enter

: accept the changes and stores the settings. 

 

To change the set point:

  

 

1)  Press and hold the (

clear

) and (

mode

) buttons for 5 seconds.  

2)  Now press the (

mode

) button nine times to display the current set point. (

RELAY

 is displayed) 

3)  Use the (

up

) and (

down

) arrow buttons to change the set point. 

4) Press 

(

enter

) to accept the change. 

5) Press 

(

clear

) to return the unit to RUN mode. (

CO

2

 ____PPM

 is displayed) 

 
CALIBRATION 

 

 The PPM-1c is factory calibrated. The sensor onboard the PPM-1c is capable of remaining in calibration for a minimum of 
three years. Extreme shock during shipping and other factors may affect the calibration. By bringing the unit outdoors, you 
can do a quick check of the calibration. The reading outdoors should be between 250-550 PPM. If the display does not read 
250-550 PPM, a simple calibration can be performed.  

 

Reviews: