SPECIFICATIONS
Selectable Temperature Range:
–4° to 140°F (–20° to 60°C) or 64° to 95°F (18° to 35°C)
Display: 7-digit LCD— Time ½ " 3-digit LCD— Temperature 9/16"
Temperature Display Update: 5 seconds
FRONT PANEL REFERENCE BUTTONS
HOUR- Advances hour digit & advances temperature in temperature setting mode
MIN- Advances minute digit
/D.R.- Hold down to set time-of-day or press to set programmed time settings to
active/inactive
TEMP.S/CLOCK- Press to toggle the display between the TEMP SET temperature
and the current temperature
- (see "Outlet Override" section)
RESET- Clears all settings
G:OFF/O:ON- Outlet Indicator light
O:ON- Orange light indicates that power at the outlet module is on.
G:OFF- Green light indicates that power at the outlet module is off.
No light indicates that the outlet module is not receiving power from the wall outlet.
DESCRIPTION OF LCD PANEL
1.
Time-of-Day controller settings
2.
TOO HI- will appear when the temperature is 5°
or more above the TEMP SET
point
TOO LO- will appear when the temperature is 5°
or below the TEMP SET point.
3.
Low battery display
4.
Time-of-day display hr/min/sec
5.
Temperature display
6.
Celsius/Fahrenheit indicator
7.
Temperature setting mode indicator
BACK PANEL REFERENCE
1.
Colder (up)/ Heater (down)
2.
F° (up)/ C° (down)
3.
Temperature Range
–20°C~60°C (up)
18°C-35°C (down)
(–4°F-140°F)
(64°F-95°F)
DIP SWITCH SETTINGS
The dip switches are located in the lower right corner on the back of the display
module. See Fig. 3 for illustrated schematic.
1.
Switch #1 selects Colder or Heater setting.
a.
Colder (up) selection turns the outlet ON when the temperature rises above
the TEMP SET point while in the temperature controlling mode. (At
temperatures below the TEMP SET point the outlet is off.)
b.
Heater (down) selection turns outlet ON when the temperature falls below
the TEMP SET point while in the temperature controlling mode. (At
temperatures above the TEMP SET the outlet is off.)
2.
Switch #2 selects °F or °C for the temperature display.
3.
Switch #3 selects the temperature range (wide or narrow see "Specifications").
TEMPERATURE CONTROLLING- HEATER MODE
Plug the electronic apparatus to be controlled into the Outlet Module. Turn on the
power switch of the electrical apparatus to be controlled. Plug the Outlet Module into a
power outlet.
1.
Set dip switch #1 to HEATER (down).
2.
Set dip switches #2 and #3 to the desired position (see "Dip Switch Settings").
3.
Press the RESET button. Wait 5 seconds for the clock and current temperature
display to reappear.
4.
Press the TEMP.S/CLOCK button, "TEMP SET" will appear in the lower right
corner of the LCD display.
5.
Press the HOUR button to advance the temperature setting to the desired
temperature.
6.
Press the TEMP.S/CLOCK button to return to the current temperature display.
7.
In order to initiate the switching function, reduce the temperature at the probe to
at least 2° below the TEMP SET point.
8.
Press the
button to trigger the outlet module on. (All subsequent ON/OFF
operations will be automatic.)
9.
Place probe in contact with the material to be measured.
The outlet will now automatically turn ON when the temperature falls 2° below the SET
TEMP point and OFF when the temperature rises above the TEMP SET point.
TEMPERATURE CONTROLLER-COLDER MODE
Plug the electronic apparatus to be controlled into the Outlet Module. Turn on the
power switch of the electrical apparatus to be controlled. Plug the Outlet Module into a
power outlet.
1.
Set dip switch #1 to COLDER (up).
2.
Set dip switches #2 and #3 to the desired position (see "Dip Switch Settings").
3.
Press the RESET button. Wait 5 seconds for the clock and current temperature
display to reappear.
4.
Press the TEMP.S/CLOCK button, "TEMP SET" will appear in the lower right
corner of the LCD display.
5.
Press the HOUR
▲
button to advance the temperature setting to the desired
temperature.
6.
Press the TEMP. S/CLOCK button to return to the current temperature display.
7.
In order to initiate the switching function, raise the temperature at the probe to at
least 2° above the TEMP SET point.
8.
Press the
button to trigger the outlet module on. (All subsequent ON/OFF
operations will be automatic.)
9.
Place probe in contact with the material to be measured.
The outlet will now automatically turn ON when the temperature rises 2° above the
SET TEMP point and OFF when the temperature falls below the TEMP SET point.
TIME-OF-DAY CONTROLLING
1.
Set dip switch #1 to "colder" (up) and switch #3 to -4°F~140°F (up).
2.
Press the RESET button. Wait 5 seconds for clock and current temperature
display to reappear.
3.
Push the TEMP.S/CLOCK button, TEMP SET appears on the display.
4.
Press the HOUR
▲
button until the display reads 0.0°F (-18°C). (TEMP SET
display defaults to 82° F. Each press of the HOUR
▲
button advances the
display. At 140° F it rolls over to –4° F.)
TRACEABLE
®
TEMPERATURE CONTROLLER INSTRUCTIONS
Traceable
®
is a registered trademark of Control Company ©2008 Control Company. 92-4130-00 Rev. 2 051908