background image

6688G 

DIGITAL MULTIMETER 

OPERATOR’S MANUAL 

1. Overview

The  multimeter  is  characterized  at  slim  size,  portable,  stable  performance  and  anti-dropping 
capacity. Using 40000 counts digit & analogue bar graph LCD monitor with character 30mm high, 
they offer clear readings. With overall circuitry design centering on large-scale IC A/D converters 
in  conjunction  and  over-load  protection  circuit,  the  meters  give  excellent  performance  and 
exquisite making as a handy utility instrument. 
The meters can be used to measure DC & AC voltage, DC & AC current, resistance, capacitor, 
frequency, duty cycle, temperature, positive diode voltage fall and audible continuity. 

2. Panel Layout

Mounting hole 

LCD display: 40000 counts digit & analogue bar graph, full function symbol display. 

SELECT

 key: This key work on the "

" range, press the key to choose resistance, 

diode  or  continuity  test,  on  the  voltage  or  current  range,  change  to  DC  or  AC,  on  the 

  range,  change  to 

  or 

  test;  If  press  and  hold  SELECT  key  to  power  on,  “Auto 

Power Off ” function will be disabled. 

RANGE

 Key: Press the “

RANGE

” key, the meter enters manual range mode, press it more 

than 2 seconds again, return to auto mode.

 

REL

  Key:  Press  the  “

REL

”  key,  the  meter  enters  relative  measuring  mode,  “

REL

”  is 

displayed on the LCD and the present reading becomes the reference value and displayed 
on the display. Relative measurement R

EL

=measurement value-Reference value

. Press it 

again to exit. 

Hz%

  Key:  On  “ACV/ACA”  or  “Hz”  range,  press  the  “

Hz%

”  key,  you  can  choose  the 

Frequency

 or 

Duty Cycle

 measurement. 

 

 

key:  Press  the  “

”  key  to  lock  display  value,  and  the  “

DH

”  sign  will 

appear on the display, press it again to exit.

 

Press the “

” key over 2 seconds

 

to 

light the back light, press it again for more than 2 seconds to lights out the back light. 

MAX/MIN

  key:  Press  the  “

MAX/MIN

”  key  to  lock 

MAX 

or 

MIN 

value,  and  the  “

MAX

”  or 

MIN

” sign will appear on the display, press it over 2 seconds to exit. 

Rotary Switch: Use this switch to select functions and ranges.

Crust of meter 

Protective casing 

 

VΩHz

 

:

 V/Hz/

  Input Jack 

 

COM:

 COM Input and Temperature “-”Jack 

 

μAmA600A

:

 μAmA, 600A current clamp adapter and Temperature “+” Input Jack 

 

20A:

 20A Input Jack 

3. Safety Information

3-1 The meter is designed according to IEC-1010 concerning electronic measuring instruments 
with an over-voltage category 1000V (CAT

and pollution 2. 

3-2 Follow all safety and operating instructions to ensure that the meter is used safely and is kept 
in good operating condition. 
3-3 safety symbols: 

Important safety information, refer to the operating manual. 

Dangerous voltage may be presence. 
Double insulation (protection Class II) 

4. Special Cautions for Operation

4-1 The meters can be safe only according to standard procedures when used in conjunctions 
with the supplied test leads. To replace damaged test leads with only the same model or same 
electric specifications. 
4-2 To avid risk of electric shock, do not use the meters before the cover is in place. 
4-3 The range switch should be right position for the testing. 
4-4  To  avoid  electric  shock  and  damaging  the  instruments,  the  input  signals  are  forbidden  to 
exceed the specified limits. 
4-5 When measuring TV set or switched power, attention should be paid to the possible pulses 
that may bring destruction to the circuit. 
4-6 Range switch position is forbidden to be changed at random during measurement. 
4-7 Take caution against shock in the course of measuring voltage higher than DC 60V & AC 30V. 
4-8 Protection fuse should be replaced only with same type and same specification. 
4-9 After operation is finished, set function switch at OFF to save battery power. 

4-10  If  the  meter  is  without  usage  for  long  time,  take  out  battery  to  avoid  damage  by  battery 
leakage. 

5. GENERAL SPECIFICATIONS

5-1 Max Voltage between input terminal and Earth Ground: CAT 

1000V

.

5-2 Over-range Indication: display “OL” for the significant digit. 
5-3 Automatic display of negative polarity “-” . 
5-4 Low Battery Indication: “

 

” displayed. 

5-5 Max LCD display: 40000 counts digit & analogue bar graph. 
5-6 Auto range & Manual range control 
5-7Auto  Power  Off:  When  measurement  exceeds 15  minutes without  switching  mode  and 

pressing key,  the  meter  will  switch  to  standby  mode.  Press any  key  to exit  standby  mode. 
When restart the system, press and hold 

SELECT

 key to disable auto power off. 

5-8 Fuse protection: 500mA/500V Fast Fuse, 20A/500V Fast Fuse 
5-9 Power supply: 9V 6F22 battery 
5-10 Operating Temp.: 0 

 

to 40 

 

(relative humidity <85%) 

5-11 Storage Temp.: -10 

 

to 50 

 

(relative humidity <85%) 

5-12 Guaranteed precision Temp.: 23±5 

 

(relative humidity <70%) 

5-13 Dimension: 218x122x75mm 
5-14 Weight: approx. 900g (including battery) 

6. Testing Specifications

Accuracy is specified for a period of year after calibration and at 18 

 

to 28 

 

(64 

 

to 82

)

 

with relative humidity to 70%. 

6-1 DC Voltage

 

Range 

Resolution 

Accuracy 

40mV 

1μV 

±(0.5% of rdg + 5 digits) 

400mV 

10μV 

±(0.1% of rdg + 2 digits) 

4V 

100μV 

40V 

1mV 

400V 

10mV 

1000V 

100mV 

±(0.1% of rdg + 5 digits) 

-- Impedance: 10MΩ, More than 100MΩ on 40mV/400mV range 
-- Overload protection: 1000V DC or 750V AC rms 

6-2 AC Voltage (True RMS) 

Range 

Resolution 

Accuracy 

40mV 

1μV 

±(1.0% of rdg + 10 digits) 

400mV 

10μV 

±(0.8% of rdg + 10 digits) 

4V 

100μV 

40V 

1mV 

400V 

10mV 

750V 

100mV 

±(1.2% of rdg + 10 digits) 

-- Impedance: 10MΩ, More than 100MΩ on 40mV/400mV range 
-- Overload protection: 1000V DC or 750V AC rms 
-- Frequency Range: 40 to 1kHz 

6-3 DC Current 

Range 

Resolution 

Accuracy 

400μA 

0.01μA 

±(1.0% of rdg + 10 digits) 

4000μA 

0.1μA 

40mA 

1μA 

400mA 

10μA 

4A 

100μA 

±(1.2% of rdg + 10 digits) 

20A 

1mA 

600A 

100mA 

±(2.5% of rdg + 10 digits) 

-- 600A range with AC/DC Current Clamp Adapter 
-- Overload protection: 500mA/500V Fast Fuse, 20A/500V Fast Fuse, 20A up to 10 seconds 

6-4 AC Current (True RMS) 

Range 

Resolution 

Accuracy 

400μA 

0.01μA 

±(1.2% of rdg + 10 digits) 

4000μA 

0.1μA 

40mA 

1μA 

400mA 

10μA 

4A 

100μA 

±(1.5% of rdg + 10 digits) 

20A 

1mA 

600A 

100mA 

±(3.0% of rdg + 10 digits) 

-- 600A range with AC/DC Current Clamp Adapter 
-- Overload protection: 500mA/500V Fast Fuse, 20A/500V Fast Fuse, 20A up to 10 seconds 
-- Frequency Range: 40 to 400Hz 

6-5 Resistance 

Range 

Resolution 

Accuracy 

400Ω 

0.01Ω 

±(1.0% of rdg + 10 digits) 

4kΩ 

0.1Ω 

±(0.5% of rdg + 10 digits) 

40kΩ 

1Ω 

400kΩ 

10Ω 

4MΩ 

100Ω 

40MΩ 

1kΩ 

±(1.0% of rdg + 10 digits) 

-- Overload protection: 500V DC or AC rms

 

6-6 Capacitance 

Range 

  Accuracy 

Resolution 

9.999nF 

±(5.0% of rdg + 10 digits) 

1pF 

99.99nF 

±(3.0% of rdg + 10 digits) 

10pF 

999.9nF 

±(2.5% of rdg + 5 digits) 

100pF 

9.999µF 

1nF 

99.99µF 

±(5.0% of rdg + 10 digits) 

10nF 

999.9µF 

±(10.0% of rdg + 20 digits) 

100nF 

9.999mF 

1µF 

99.99mF 

10µF 

-- Overload protection: 500V DC or AC rms

 

Reviews: