background image

TB-3043 Page 1 of 5

© 2009 DESCO INDUSTRIES, INC.

Employee Owned

Figure 1. 

60505

 High Output Benchtop Ionizer

Description

The Desco 

60505

, High Output Benchtop Ionizer is a steady 

state DC, auto-balancing tabletop ionizer. A specially designed 

fan module creates better coverage (16” x 36”), consistent offset 

voltage (balance) ±3 should be with a discharge times at less than 

one second at 12”.  The unit continually adjusts voltage output for 

optimal performance and automatically alarms when maintenance 

is needed or the unit is not operating correctly.  Ionization offset 

voltage balance is consistent at all 3 speeds (50-100 CFM) and 

allows for use in sensitive applications, such as calibration, test, or 

repair procedures.  The powder coated housing, multi-mount stand 

and sealed bearing fans minimize foreign objects debris, corrosion, 

and other contaminations when used in clean or or sensitive 

areas.  Field replaceable emitter pins allow for replacement without 

returning to the factory.

“Necessary non-conductors in the environment cannot lose 

their electrostatic charge by attachment to ground. Ionization 

systems provide neutralization of charges on these necessary 

non-conductive items (circuit board materials and some 

device packages are examples of necessary non-conductors).  

Assessment of the ESD hazard created by electrostatic charges 

on the necessary nonconductors in the work place is required to 

ensure that appropriate actions are implemented, commensurate 

with risk to ESDS [ESD sensitive] items”. (ANSI/ESD S20.20-2007 

Foreword)

“In order to mitigate field-induced CDM [Charged Device Model] 

damage, the ESD program shall include a plan for the handling of 

process-required insulators.  If the field exceeds 2,000 volts/inch, 

steps shall be taken to either: A) Separate the insulator from the 

ESD-sensitive device by a distance of 30 cm (12 inches); or B) Use 

ionization or other charge mitigating techniques to neutralize the 

charge.” (ANSI/ESD S20.20-2007 section 8.3)

“The primary method of static charge control is direct connection to 

ground for conductors, static dissipative materials, and personnel.  

A complete static control program must also deal with isolated 

conductors that cannot be grounded, insulating materials (e.g., 

most common plastics), and moving personnel who cannot use 

wrist or heel straps or ESD control flooring and footwear.  Air 

ionization is not a replacement for grounding methods.  It is one 

component of a complete static control program.  Ionizers are used 

when it is not possible to properly ground everything and as backup 

to other static control methods.  In clean rooms, air ionization 

may be one of the few methods of static control available.” (ESD 

Handbook ESD TR20.20 Ionization, section 5.3.6.1)

“All ionization devices will require periodic maintenance for 

proper operation.  Maintenance intervals for ionizers vary 

widely depending on the type of ionization equipment and use 

environment.  Critical clean room uses will generally require more 

frequent attention.  It is important to set-up a routine schedule for 

ionizer service.  Routine service is typically required to meet quality 

audit requirements.” (ESD Handbook ESD TR20.20 section 5.3.6.7 

Maintenance / Cleaning)

CSA International tests products for compliance to national and 

international standards, and issues certification marks for qualified 

products.  The certification marks tell potential customers and users 

that a product has been evaluated by a formal process-involving 

examination, testing and follow-up inspection-and that it complies 

with applicable standards for safety and performance.

http://www.csa-international.org/

High Output

1 Benchtop Ionizer with stand

1 Power Cord

1 Emitter Point Cleaner Pack (Item 

60506

)

1 Certificate of Calibration

Input Voltage

WARNING - RISK OF ELECTRIC SHOCK

THESE SERVICING INSTRUCTIONS ARE FOR USE BY 

QUALIFIED PERSONNEL ONLY.  DO NOT PERFORM ANY 

SERVICING OF INTERNAL PARTS UNLESS YOU ARE 

QUALIFIED TO DO SO.

NOTE:  THE AC POWER CORD MUST ALWAYS BE 

DISCONNECTED BEFORE THE UNIT IS DISASSEMBLED.  UNIT 

SHIPPED SET AT 110V

High Output Benchtop Ionizer
Installation, Operation and Maintenance

TECHNICAL BULLETIN TB-3043

TB-3043 August 2010 Page 1 of 5

© 2010 DESCO INDUSTRIES INC.

Employee Owned

Made in America

Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com 

Summary of Contents for 60505

Page 1: ...c dissipative materials and personnel A complete static control program must also deal with isolated conductors that cannot be grounded insulating materials e g most common plastics and moving personnel who cannot use wrist or heel straps or ESD control flooring and footwear Air ionization is not a replacement for grounding methods It is one component of a complete static control program Ionizers ...

Page 2: ...t will conduct a self test The audible alarm will sound and then the LED will cycle through RED YELLOW and GREEN The LED will remain GREEN during normal operation Figure 4 Front and Back High Output Benchtop Ionizer Angle Adjustment Knobs The angle of the fan may be adjusted to direct the airflow in the optimum coverage area Visual Alarm GREEN Normal Operation RED Ionizer needs maintence Voltage s...

Page 3: ...ntification method for each ionizer in your Maintence Calibration schedule You may choose to use the ionizer s serial number for identification as it is a unique number for identification as it is a unique number that can be traced to when the ionizer was manufactured Neutralization Discharge Times The comparative efficiency of overhead ionizers is determined by a standard test published by ANSI E...

Page 4: ...ge time testing If the unit does not meet offset voltage specifications or minimum established discharge time limits further service is indicated Manufacturers should provide details on service procedures and typical service intervals JESDD625 A Revision of EIA 625 recommends ionizers be tested semiannually noting to use S3 1 except the number of measurement points and locations may be selected ba...

Page 5: ...ng instructions and ship your unit freight prepaid Warranty Exclusions THE FOREGOING EXPRESS WARRANTY IS MADE IN LIEU OF ALL OTHER PRODUCT WARRANTIES EXPRESSED AND IMPLIED INCLUDING MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE WHICH ARE SPECIFICALLY DISCLAIMED The express warranty will not apply to defects or damage due to accidents neglect misuse alterations operator error or failure to p...

Reviews: