1.

  Install version 17.1.6 or higher of DicksonWare Software  

 

2.

  Unscrew the logger cover; connect the cable (supplied with the software) to the logger and to a working USB port on your 

computer.  

 

3.

  Click the Setup button. The Setup window will appear, and all fields should be automatically filled in, this will confirm that 

DicksonWare has recognized the logger. Should all fields remain blank, refer to “No Communication” in the Support section 
at www.dicksondata.com for this logger model. Once DicksonWare recognizes the logger, press the 

CLEAR

 button. This will 

delete all data currently stored.  

 

4

.  The logger is now sampling and ready for use. The default sample interval is 5 minutes. You may choose to change the 

default setting for sample interval and start date and time. When changing these settings. The logger will automatically go 
through the 

CLEAR

 process to accept your changes.  

NOTE

: To conserve battery life when storing the unit, press the 

“Setup” button and select “Stop When Full” 

 

WARNING

: Fire, explosion, and severe burn hazard.  Do not recharge, disassemble, heat above 85°C, or incinerate or expose 

internal battery to water 

 
 

Useful Features 

 

1.

  

Delayed Start

: Program your logger to the date/tim Delayed Start: e to begin logging.  

 

2.

  

Real Time Monitoring

: Monitor and graph your proce Real Time Monitoring: ss as it happens and print and save from real 

time screen.  

 

3.

  

Customize Graphed Data

: DicksonWare calculates Min Customize Graphed Data: , Max and Average of all data collected. 

Customize data by highlighting specific data points which will recalculate MIN, MAX and Average for highlighted information.  

 

4.

  

Exported Data

: A snapshot of your graph or a real t Exported Data: ime graph of points can be easily exported to other 

programs such as Excel or PowerPoint.  

 
Click on the 

FEATURES

 button in the software for detailed instructions on how to use above features. 

 