Getting Started

Dickson SP425 & TP425

Start/Stop Data Logger with Display Operation

  Install DicksonWare™ (compatible with Windows® 

98 or higher). DicksonWare™ version 14.8 or higher 

required. 

  If you already have DicksonWare on your PC check 

the version by choosing “Help” and “About” from the 

menu bar to check what version you have and make 

sure it complies with the requirements of the logger.

 Open DicksonWare™ using the icon on your desktop.
  Connect the cable (supplied with the software) to the 

logger and to a working USB port on your computer.

  Click the Setup button. You will be prompted to 

select USB or Serial COM port, select USB and click 

Continue. A Setup window will appear, and 

all fields should be automatically filled in. This 

confirms that DicksonWare™ has recognized the 

logger. The logger is now sampling.

Display 

Below are the possible display states and the conditions 

they represent:
Solid or Scrolling (Logging)

  Logging normally. Pressing Start/Stop button will  

stop logging. Pressing Start/Stop button again will 

restart logging.

   

No Delay

 selected in Setup–Logging normally 

Flashing (Not Logging)

  

Start after Delay

 selected in Setup but logging has 

not yet begun due to delay.


  The logger is full and Stop When Full has been 

selected in Setup. To clear logger, download to 

DicksonWare and select Clear.

Blank (Not Logging)

  Press Button to Start has been selected in Setup but 

the Start/Stop button has not been pressed.

  Start/Stop button has been pressed displaying “OF” 

(short for OFF) to stop logging.

“OF”

  Displays for two seconds after Start/Stop button is 

pressed to stop logging.

“ON”

  Displays for two seconds after Start/Stop button is 

pressed to start logging.

Min/Max

There are two display options: Current Readings only or 

Current/Min/Max scroll, which can be selected under 

Setup/Display.
To reset Min/Max readings on the display, a paper clip or 

toothpick should be used to depress the Min/Max Clear 

button located on the back of the unit. This will not clear 

logged data.

Press Button to Start

This feature needs to be selected when the logger is  

setup (Setup/Samples). Hold down the Start/Stop button 

on the front of the unit until “ON” appears on the display. 

The logger is now logging. Pressing the button again will 

stop logging.

Need to Know

The Dickson Warranty

 When changing logger settings (sample interval and 

start date and time) the logger will automatically clear 

all stored data.
 To achieve longer battery life during operation, use a 

less frequent sample rate and disconnect the unit from 

the USB or serial port when not downloading data.

Fahrenheit/Celsius

The data logger is defaulted to log data in Fahrenheit. If 

you want to view future logged data on the graph or table 

in Celsius go to “File,” “Preferences” and then change the 

temperature selection. 

Battery Replacement

The battery level monitor on the “Setup” screen displays 

the battery voltage and a low battery warning when 

replacement is needed.

When changing the battery the logger will not collect 

data however memory will not be lost. To start sampling 

again, download the data and then clear the memory.

Setting Plot Style

Spline and Points & Spline should not be used as they will 

auto fill the graph over the period(s) where the logger is 

off and not logging.

Start Mode

This logger has three start modes:

1.  No Delay
2  Press Button to Start
3  Start After Delay

Each time the logger is cleared it will default to No Delay 

and start logging immediately. To change start mode clock 

on the Start Date/Time Setup button and select Delay 

Type. Click ok.

Dickson warrants this line of instrument will be free from 

defects in material and workmanship under normal use 

and service for a period of twelve months after delivery.

This warranty does not cover routine calibration and 

battery replacement
For Specifications and Technical Support go to  

www.DicksonData.com

.