background image

dIXEL

Operating and Instructions Manual

1592006100

1592006100  XB570L GB r1.0 04.11.2004.doc

XB570L

1/6

XB570L

BLAST CHILLER  - QUICK CHILL AND HOLD FUNCTION

Contents

1.

General warning _______________________________________ 1

2.

General Features ______________________________________ 1

3.

Mounting & Installation __________________________________ 1

4.

Electrical Connections __________________________________ 1

5.

Connections __________________________________________ 1

6.

Frontal panel _________________________________________ 1

7.

Front Panel Commands _________________________________ 2

8.

How To Select A Cycle. _________________________________ 2

9.

How To Start The Selected Cycle. _________________________ 2

10. How to temporarily stop the running cycle. __________________ 2
11. How to definitively stop the cycle. _________________________ 2
12. How To Modify The Set Point Of The Hold Mode. _____________ 2
13. How to start a manual defrost. ____________________________ 2
14. How A Cycle Is Done. __________________________________ 2
15. Function And Parameter Programming Menu ________________ 3
16. Accessing “Pr2” and SECURITY CODE input ________________ 4
17. List of parameters______________________________________ 4
18. Printer management____________________________________ 5
19. IV configurable relay. ___________________________________ 5
20. Digital configurable input. ________________________________ 5
21. ALARM SIGNALS _____________________________________ 5
22. Technical data ________________________________________ 6
23. Standard Value of the cycles._____________________________ 6
24. Standard Values of the parameters.________________________ 6

1. General 

warning

1.1 

 PLEASE READ BEFORE USING THIS MANUAL

  This manual is part of the product and should be kept near the

instrument for easy and quick reference.

  The instrument shall not be used for purposes different from those

described hereunder. It cannot be used as a safety device.

 

Check the application limits before proceeding.

1.2 

 SAFETY PRECAUTIONS

 

Check the supply voltage is correct before connecting the instrument.

 

Do not expose to water or moisture: use the controller only within the
operating limits avoiding sudden temperature changes with high
atmospheric humidity to prevent formation of condensation

  Warning: disconnect all electrical connections before any kind of

maintenance.

 

The instrument must not be opened.

 

In case of failure or faulty operation send the instrument back to the
distributor or to “Dixell s.r.l.” (see address) with a detailed description of
the fault.

 

Consider the maximum current which can be applied to each relay (see
Technical Data).

 

Ensure that the wires for probes, loads and the power supply are separated

and far enough from each other, without crossing or intertwining.

 

Fit the probe where it is not accessible by the end user.

 

In case of applications in industrial environments, the use of mains filters
(our mod. FT1) in parallel with inductive loads could be useful.

2. General 

Features

The series XB has been created for fast chilling or freezing goods according
to international food safety standards.

There are FOUR types of cycles:

 

The CYCLES: Cy1, Cy2, Cy3, Cy4 are pre-set according to the most
common cycles used in food - safety applications; the user can select
one of them according to his own requirements and modify it as he
wants.

 

Any cycle can be manually terminated before the normal.

 

Any cycle can use the third probe like an “insert probe”, it measures the
internal temperature of the product.

 

During the Cycle there are no defrosts and the fans are always on, a
defrost cycle can be done before any freezing cycle.

 

The cycle is divided up to 3 phases completely configurable by the user.

 

Each instrument is provided with an output for remote display XR REP,
which shows the temperature of cabinets or goods.

 

The XB570L controller is provided with internal real time clock and can
be connected to the XB05PR printer. This means that  a report, which

includes all the main features of cycle, can be printed: start and end of
the cycle, length of the cycle, logging of the temperature of the cabinet
and goods.

3.  Mounting & Installation

Model XB570L is a controller panel mounted, hole dims 150x31 mm, and
fixed with the screws. The ambient working temperature range should be
between 0÷60°C. Avoid locations subject to heavy vibration, corrosive gases
or excessive dirt. The same applies to the probes. Ensure ventilation around
the instrument.

4. Electrical 

Connections

The instruments are provided with a screw terminal block to connect cables

with a cross section up to 2,5 mm

2

. Before connecting cables make sure the

power supply complies with the instrument’s requirements. Separate the
input connection cables from the power supply cables, from the outputs and
the power connections. 

Do not exceed the maximum current allowed on

each relay

, in case of heavier loads use a suitable external relay.

4.1 PROBES 

CONNECTION

The probes shall be mounted with the bulb upwards to prevent damages due
to casual liquid infiltration. It is recommended to place the thermostat probe
away from air streams to correctly measure the average room temperature.
Place the defrost termination probe among the evaporator fins in the coldest
place, where most ice is formed, far from heaters and from the warmest
place during defrost, to prevent premature defrost termination.

5. Connections

Di

gi

ta

l i

np.

 

Ev

ap.

Ins

er

t

Room

XB570L

Def.

Alarm/

AUX

Fan

12A

250Vac

MAX

20A

8A

250Vac

8A

250Vac

Supply

230Vac

Comp

8A

250Vac

19

13 14 15 16

12

5 6 7 8 9 10 11

1 2 3 4

17 18

6. Frontal 

panel

Aux

1

2

3

4

6.1  MEANING OF THE LEDS’

A series of light points on the front panels is used to monitor the loads
controlled by the instrument. Each LED function is described in the following
table.

LED

MODE

ACTION

ON

- Compressor enabled

Flashing

- Programming Phase (flashing with LED 

)

- Anti-short cycle delay enabled

ON

- Fans enabled

Flashing

- Programming Phase (flashing with LED 

)

- Activation delay active

ON

- Defrost active

Flashing

- Drip time active

ON

- Freezing cycle or hold mode active

Flashing

- Instrument temporarily stop

ON

- Alarm signalling
- In “

Pr2

” indicates the parameter is also present

in “Pr1”

AUX

ON

- IV relay enabled

Summary of Contents for XB570L

Page 1: ...tures The series XB has been created for fast chilling or freezing goods according to international food safety standards There are FOUR types of cycles The CYCLES Cy1 Cy2 Cy3 Cy4 are pre set according to the most common cycles used in food safety applications the user can select one of them according to his own requirements and modify it as he wants Any cycle can be manually terminated before the...

Page 2: ...ys C1 C2 C3 C4 the related led will be lighted and the cycle will be selected NOTE to pass from a cycle to another one simply push the related key HOLD PHASE To select the old phase deselect the enabled cycle pushing the related key ES Cycle 1 is selected its led is oon push the C1 key the led will be turn off and the hold phase will be enabled 2 Cycles are pre set with the following values 1 Cy1 ...

Page 3: ...eaches the iS3 value normally 18 C 0 F in any case it ends when the maximum time Pd1 Pd2 Pd3 has expired 14 3 4 End of the Blast Chill cycle and starting of the Hold Mode When the Blast Chiller cycle ends an alarm signal is generated buzzer and alarm relay is turned ON the display shows the message End alternating with the room temperature The alarm automatically stops after 30min or by pressing a...

Page 4: ...sor ON time with faulty probe 0 255 min time during which the compressor is active in case of faulty thermostat probe With COn 0 compressor is always OFF COF Compressor OFF time with faulty probe 0 255 min time during which the compressor is off in case of faulty thermostat probe With COF 0 compressor is always active diC Digital input operating mode EAL bAL dor EAL external alarm bAL serious alar...

Page 5: ...the DOWN key or by closing the door 19 1 3 OAC UL UV light for a sterilisation cycle It is used for a sterilisation cycle of the machine It can be enabled ONLY WHEN THE CONTROLLER IS NOT WORKING or when a manual defrost is running The ultra violet light stays on for the OAt time It can be switched off before the OAt time by pressing the DOWN key NOTE with OAt 0 it can be switched off only by press...

Page 6: ... rS2 30 C 22 F Pd3 OFF rS1 30 C 22 F Pd2 OFF dbH yes Pd1 4 0 iS3 18 C 0 F HdS 18 C 0 F Cy4 direct fast freezing without holding dbC no iS2 18 C 0 F rS3 30 C 22 F iS1 18 C 0 F rS2 30 C 22 F Pd3 OFF rS1 30 C 22 F Pd2 OFF dbH no Pd1 4 0 iS3 18 C 0 F HdS OFF 24 Standard Values of the parameters Lab Description XB570L C F Regulation Hy differential 2 4 AC Anti short cycle delay 1 1c2 Second compressor ...

Reviews: