OPERATING INSTRUCTIONS
SERIES MM10 DIGITAL MULTIMETER
SAFETY INFORMATION
The following safety information must be observed to insure
maximum personal safety during the operation of this meter.
Do not use the meter if the meter or test leads look damaged, or
if you suspect that the meter is not operating properly.
Never ground yourself when taking electrical measurements. Do
not touch exposed metal pipes, outlets, fixtures, etc., which might
be a ground potential. Keep your body isolated from ground by
using dry clothing, rubber shoes, rubber mats, or any approved
insulating material.
Turn off power to the circuit under test before cutting,
unsoldering, or breaking the circuit. Small amounts of current
can be dangerous.
Use caution when working above DC 60V or AC 30Vrms. Such
voltages pose a shock hazard.
When using the probes, keep your fingers behind the finger
guards on the probes.
Measuring voltage which exceeds the limits of the multimeter
may damage the meter and expose the operator to a shock
hazard. Always recognize the meter voltage limits as stated on
the front of the meter.
GENERAL SPECIFICATIONS
Display:
3-1/2 digit liquid crystal display (LCD) with a maximum
reading of 1999.
Polarity:
Automatic, positive implied, and (-) for negative polarity
indication.
Overload:
(1) or (-1) is displayed on LCD.
Low Battery Indication:
When "
" shows on LCD, battery
needs to be replaced.
Sampling Rate:
2.5 times per second.
Operating Environment:
0°C~40°C 80% RH.
Storage Temperature:
-10°C~60°C 0~80% RH with battery
removed from meter.
Power:
One 006P 9V standard battery.
Battery Life:
150 hours approx.
Dimension:
125 mm (H) x 68 mm (W) x 25 mm (D).
Weight:
150g approx. including battery.
Accessory:
Instruction manual, test leads, (1) 9V battery.
SPECIFICATIONS
DCV
Range Resolution Accuracy
Input
Impedance
Overload
Protection
2V
1mV
0.5% + 1 dgt
20V 10mV
200V 100mV
600V 1V
1.5% + 1 dgt
10M
Ω
DC 600V
AC 600V
rms
ACV
Range Resolution
Accuracy
Input
Impedance
Overload
Protection
200V 100mV
2% + 3 dgt
(40 Hz-
500 Hz)
600V
1V
2.5% + 3 dgt
4.5M
Ω
DC 600V
AC 600V
rms
DCA
Range Resolution
Accuracy
Input
Impedance
Overload
Protection
2mA 1
μ
A
20mA 10
μ
A
200mA 100
μ
A
1.5% + 1 dgt
0.3V
0.5A Fuse
& Diodes
10A
10mA
2% + 1 dgt
0.7V
Resistance (
Ω
)
Range Resolution
Accuracy
Open
Voltage
Overload
Protection
200
Ω
0.1
Ω
2% + 2 dgt
2K
Ω
1
Ω
2.8V
20K
Ω
10
Ω
200K
Ω
100
Ω
2M
Ω
1K
Ω
1.5% + 2 dgt
20M
Ω
10K
Ω
2.5% + 2 dgt
0.3V
DC/AC
500V
rms
Diode Test
Range Resolution
Test
Current
Open
Voltage
Overload
Protection
1mV 1mA
≈
2.8V
DC/AC
500V rms
Continuity Test
x
Range Resolution Sound
Open
Voltage
Overload
Protection
x
0.1
Ω
<30
Ω
<
≈
2.8V
AC/DC
500V rms
Transistor hFE
Range
Measuring
Scope
Test
Voltage
Test
Current
Overload
Protection
NPN
PNP
0-1000 <3.2V IB
≈
10
μ
A
LED
Range
Measuring
Scope
Test
Voltage
Test
Current
Overload
Protection
LED 1.5-3V 3V
≈
25mA