OPERATING INSTRUCTIONS 

 

SERIES MM10 DIGITAL MULTIMETER 

 
 

 
 

SAFETY INFORMATION 

The following safety information must be observed to insure 
maximum personal safety during the operation of this meter. 
 
Do not use the meter if the meter or test leads look damaged, or 
if you suspect that the meter is not operating properly.   
 
Never ground yourself when taking electrical measurements.  Do 
not touch exposed metal pipes, outlets, fixtures, etc., which might 
be a ground potential.  Keep your body isolated from ground by 
using dry clothing, rubber shoes, rubber mats, or any approved 
insulating material. 
 
Turn off power to the circuit under test before cutting, 
unsoldering, or breaking the circuit.  Small amounts of current 
can be dangerous. 
 
Use caution when working above DC 60V or AC 30Vrms.  Such 
voltages pose a shock hazard. 
 
When using the probes, keep your fingers behind the finger 
guards on the probes. 
 
Measuring voltage which exceeds the limits of the multimeter 
may damage the meter and expose the operator to a shock 
hazard.  Always recognize the meter voltage limits as stated on 
the front of the meter. 

 
 

GENERAL SPECIFICATIONS 

Display:

 3-1/2 digit liquid crystal display (LCD) with a maximum 

reading of 1999. 

Polarity:

 Automatic, positive implied, and (-) for negative polarity 

indication. 

Overload:

 (1) or (-1) is displayed on LCD. 

Low Battery Indication: 

 When " 

 " shows on LCD, battery 

needs to be replaced. 

Sampling Rate:

 2.5 times per second. 

Operating Environment:

 0°C~40°C 80% RH. 

Storage Temperature:

 -10°C~60°C 0~80% RH with battery 

removed from meter. 

Power:

 One 006P 9V standard battery. 

Battery Life:

 150 hours approx. 

Dimension:

 125 mm (H) x 68 mm (W) x 25 mm (D). 

Weight:

 150g approx. including battery. 

Accessory: 

Instruction manual, test leads, (1) 9V battery. 

 
 

SPECIFICATIONS 

DCV 

Range Resolution  Accuracy 

Input 

Impedance 

Overload 

Protection 

2V 

1mV 

0.5% + 1 dgt 

20V 10mV 

200V 100mV 
600V 1V 

1.5% + 1 dgt 

10M

 

DC 600V 

AC 600V 

rms 

 

ACV 

Range Resolution

Accuracy 

Input 

Impedance

Overload 

Protection 

200V 100mV 

2% + 3 dgt 

(40 Hz- 

500 Hz) 

600V 

1V 

2.5% + 3 dgt 

4.5M

 

DC 600V 

AC 600V 

rms 

 

DCA 

Range Resolution

Accuracy 

Input 

Impedance

Overload 

Protection 

2mA 1

μ

20mA 10

μ

200mA 100

μ

1.5% + 1 dgt 

0.3V 

0.5A Fuse 

& Diodes 

10A 

10mA 

2% + 1 dgt 

0.7V 

 

 

Resistance (

Range Resolution

Accuracy 

Open 

Voltage 

Overload 

Protection 

200

 0.1

 

2% + 2 dgt 

2K

 

1

 

2.8V 

20K

 10

 

200K

 100

 

2M

 1K

 

1.5% + 2 dgt 

20M

 10K

 

2.5% + 2 dgt 

0.3V 

DC/AC 

500V 

rms 

 

Diode Test 

Range Resolution 

Test 

Current 

Open 

Voltage 

Overload 

Protection 

 

1mV 1mA 

2.8V 

DC/AC 

500V rms 

 

Continuity Test 

x

 

Range Resolution  Sound 

Open 

Voltage 

Overload 

Protection 

x

 

0.1

 <30

 

<

2.8V 

AC/DC 

500V rms 

 

Transistor hFE 

Range 

Measuring 

Scope 

Test 

Voltage 

Test 

Current 

Overload 

Protection 

NPN 
PNP 

0-1000 <3.2V IB

10

μ

A

 

 

 

LED 

Range 

Measuring 

Scope 

Test 

Voltage 

Test 

Current 

Overload 

Protection 

LED 1.5-3V  3V 

25mA  

 