background image

INTENDED USE

The Control Solution is a desired aqueous glucose solution with small amounts of non-reactive ingredients and food coloring for verifying 
the accuracy of the entire Blood Glucose Monitoring System. The Control Solution test examines the performance of the Glucose Meter, 
Test Strip, and your operating techniques.

WHEN TO PERFORM A CONTROL SOLUTION TEST

EasyTouch® Control Solution can be used to verify that the EasyTouch® Glucose Monitoring system is working correctly and giving 
accurate results.  The system is performing correctly if the control solution test result falls within the specific control solution range listed 
on your Test Strip vial. You are encouraged to use Control Solution on a regular basis.  
If you suspect your blood glucose test are not consistent with how you feel or your test are higher or lower than normal, you may validate 
the performance of your test with Control Solution.

CONTROL SOLUTION CONTENTS

Each Control Solution is provided as a buffered, stabilized, 
aqueous solution of desired D(+)-glucose containing a 
preservative and other non-reactive ingredients. The solutions 
are formulated to react equivalently to a whole blood 
sample containing approximate 50 or 300 mg/dL glucose 
concentrations. Compare control solution test results against 
the ranges printed on your Glucose Test Strip vial.

CAUTION: Control Solution is for in vitro use only. It is not to be 
swallowed or injected. The use of the Control Solution is 
recommended to verify proper system performance and should 
not be considered as a substitution of actual capillary whole blood 
testing for blood glucose level.

MATERIALS REQUIRED

1. Glucose Meter
2. Test Strip
3. Control Solution
4. Clean, nonabsorbent surface

Note: Please make sure you are using the same model of Test Strip and Glucose Meter.

TEST PROCEDURE

PREPARATION

After you have inserted a test strip into your Glucose Meter, make sure that you have set the Glucose Meter in the Control Solution test 
“ctl“ mode, so the result will not be stored in the memory and not be incorporated into the average. Please read your Blood Glucose 
Monitoring System User Manual. 

Easy

Touch

®  

Control Solution Instructions

WARNING: 

Re-cap immediately & 

always close lid tightly.

Store at 39-86°F 

(4-30°C). 

Do not refrigerate. 

Always store test strips inside vial.

Discard Date: ______________

Discard 3 months after opening.

Made in Taiwan. Distributed by:

MHC Medical Products, LLC 

8695 Seward Rd, Fairfield, OH 45011

AD-DH-042221.7  •  Rev date: 04/2021

For testing of glucose in capillary whole blood.

For in vitro diagnostic use. 

EXP

50

 

TEST STRIPS

Glucose Test Strips

For use with Easy•Touch® Meter.

Easy•Touch

®

  

mg/dL 

mmol/L

Low:  35–65 1.9–3.6
High:  263–353 14.6–19.6

Example 

Low (Level 1) Reading

Example 

High (Level 3) Reading