WEIGHT AND FREQUENCY  CALIBRATION PROCEDURE FOR PLUS/SE20 
2000g  SCALE

 

1.  Remove both case screws underneath the keyboard in front of the unit. 
2.  Use a coin to pop open the case  in front of the unit to allow a small blade screw driver to move in. 

Or underneath the scale there two slots which will fit a mini-screw driver for calibration job. 

3. 

Plug in the supplied AC adapter to power supply or put a “9V” Alkaline battery and the scale is 
ready for operation.  

 

4. 

Press the key “ON/OFF”  follows by “TARE”  and  holds momentarily.

 

5. 

when display shows 

 - - - -

 (minus sign) immediately  press the  “ON/OFF”  key to enter the 

calibration mode routine.

 THE SCALE SHALL SHOW  E.E  or  -80  [

USE ONLY THE SUPPLIED 

ADAPTER!!!

 

Notes:- 
If the scale shows calibration mode within this range  as below:-  
a) WITH PLATFORM MOUNTED,    200< CALIBRATION MODE< 240 or  
b) WITHOUT PLATFORM MOUNTED,  90< CALIBRATION MODE < 140; 
then the scale does not need any zero adjustment or offset frequency adjustment 

 
 To adjust the zero offset (claibration mode)  to read within  the range of . 200  to . 240 
 ( with platform) by turning potentiometer of VAR2.  (Refer to figure below) 

 
(If more than > 240 the scale calibration mode 
 need to be adjusted this way and  vise versa)                                                                                                              

Legends:-  
VAR1 = 500 ohm (501) multiturn variable resistor to adjust weight. 
VAR2 = 10 Kohm (103)  multiturn variable resistor to adjust zero or calibration mode. 

 
5)  To increase the frequency ( 

if scale shows wait -80

), adjust the variable resistor in the ANTI-

CLOCKWISE direction.  To decrease the frequency (

if scale shows E.E

), adjust the VAR2 

(variable resistor) in the CLOCKWISE direction. Please turn this VAR2 untill the reading on the 
LCD shows the desired range. 

 
 
 
6)  If the calibration mode on LCD display shows:-  
“.  240”, then place a standard weight of say 2000g  weight on the scale’s platform and adjust now the 
span by turning VAR1 to read (240+2000= .2240) on the LCD display.  (refer to above figure for which 
potentiometer to be used)  Or use any other standard  weight and adjust accordingly.  
7)  Remove the standard weight now, and read the calibration mode again, if it is not “.  240, please  

repeat the adjustment on VAR2 again, please turn the VAR2 until the LCD display shows “.  240” 