V
V
I
I
P
P
C
C
L
L
A
A
M
M
P
P
M
M
I
I
N
N
I
I
PRECAUTIONS FOR USE
Never use on networks with voltage higher than 600 V
in relation to the ground and whose overvoltage category
is higher than III, i.e. fixed industrial and domestic
installations (see IEC 664-1).
Indoor use in environments with a degree of pollution
of 2 maximum (cf. IEC 664-1), temperature of 0°C to +
50°C and relative humidity lower than 70%.
Use accessories compliant with safety standards
(NF EN 61010-2-031) with minimal voltage of 600 V
and overvoltage category III.
Never open the clamp box before disconnecting all
power sources.
Never connect to the circuit to be measured if the clamp
box is not properly closed.
Before any measurement, check the proper positioning
of the cables and switch.
When measuring current check for proper alignement
of the conductor in relation to the markers and proper
closing of the jaws.
Always disconnect the clamp from any power source
before changing the battery.
Do not perform resistance tests, continuity tests or semi-
conductor tests on a circuit under power.
USER MANUAL
Thank you for purchasing this
VIP CLAMP
MINI series multimeter clamp
.
To get the best service from this instrument:
_
read
this users manual carefully,
_
respect
the safety precautions detailed
Via Vizzano,44 - 40037 Pontecchio Marconi (Bologna) Italy
Tel.+39 0516782006 Fax +39 051845544
......................................................................................
http://www.elcontrol-energy.net
e-mail (Comm.le Italia) vendite@elcontrol-energy.net
e-mail (Sales Dept.) sales@elcontrol-energy.net
English