background image

1. Installation

  DPT-D series actuators are high precision 

instruments with sub-nanometre resolution 

and stability. They require careful handling 

and must not be dropped or subjected to 

unnecessary  torque on installation 

(see 

section 3)

.

 

1.1

 

Mounting into a system

 

  The DPT-D series of actuators can be 

mounted by means of an M3 screw 

requiring 3mm maximum of thread 

engagement. A pair of Ø1.6mm dowels 

requiring 2mm maximum of engagement 

as shown in Figure 1 can be used for 

accurate location if required. Mounting 

surface for the actuator base should be flat 

to better than 5µm. Full dimensional details 

are available on the Installation drawing 

PA0099-IN-001. 

queensgate

a brand of Elektron Technology

Queensgate Instruments 

info@queensgate.com 
www.nanopositioning.com 

 

  Closed loop positioner

 DPT-D

  Installation and Handling Instructions

2. Operation

  Connect the actuator to the corresponding 

controller socket, ensure the screw lock 

posts are secured to a torque of 25-30cNm.

  Avoid prolonged use in a high humidity 

environment and extended use at high 

voltage as this may lead to an increase 

in leakage current which in turn will 

dramatically reduce actuator lifespan. 

  Use the DPT-D actuators only in 

conjunction with original calibrated

 

1.2 Actuator Connector

  Keep the EMC cap fitted to the actuator 

connector until ready to connect to the 

controller. Ensure the controller is switched 

off before installing or removing cables.

  Static discharge to the connector may 

cause damage, ensure that all personnel 

handling the unit are adequately grounded.

3. Attaching or changing external  
  load to the head

  When attaching an external load or 

accessory to the actuator head always use 

the spanner supplied. Failure to comply 

may result in damage. The maximum 

tightening torque applied must not exceed 

50cNm. There is no anti-rotation feature on 

the actuator head: If head rotates damage 

will occur. 

4. Maintenance

  Do not contaminate the surface of the 

assembly by exposing it to conductive or 

corrosive substances. It is recommended 

that isopropanol is used when cleaning the 

assembly. Do not use acetone. 

Closed-loop preloaded digital piezo translators with very high linearity and repeatability 

5. Accessories

  VEP3: V-Groove End Piece

  FS25-1”: 25mm diameter Mirror Holder 

  FS12-1/2”:  12.5mm diameter Mirror Holder 

  BEP5: Spherical End Piece

  PEP: Plain End Piece

  MEP: Magnetic End Piece

  MBI-D: Mounting Block

DPT-D-

020

DPT-D-

060

DPT-D-

040

DPT-D-

010

DPT-D-

100

DPT-D-

120

DPT-D-

080

  Mounting the DPT-D on the housing is not 

recommended and may be detrimental to 

hysteresis and thermal drift. For optimum 

performance mount the DPT-D vertically. 

Apply only axial forces to the actuator head 

referring to the DPT-D datasheet, bending 

or torsion forces should be avoided. 

Figure 1.

  Queensgate controllers and accessories. 

For further information on system 

installation and use refer to the controller 

user manual

.

Reviews: