background image

 

 

 

    

Page 1 

 

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 

 

euromex microscopen bv 
The Netherlands 

www.euromex.com

  

Summary of Contents for iScope IS.1152-EPL

Page 1: ...Page 1 euromex microscopen bv The Netherlands www euromex com ...

Page 2: ... production methods resulted also in an excellent price performance ratio Models Objectives 1 optional objectives 2 oil immersion objectives Models Bino Trino EWF 10x 20 mm EWF 10x 22 mm E Plan objectives Plan Phase objectives E Plan IOS objectives Plan IOS pbjectives Plan Phase IOS objectives Mechanical rackless stage iCare sensor Köhler LED IS 1152 EPL IS 1153 EPL IS 1152 PLPH IS 1153 PLPH IS 11...

Page 3: ... the objective is an indication for the resolving power of the objective The total magnification can be calculated by multiplying the magnification of the eyepiece with the magnification of the objective The magnifications are displayed in the table below Eyepiece Objective Magnification 10x 4x 40x 10x 10x 100x 10x 40x 400x 10x 60x 600x 10x 100x 1000x ...

Page 4: ...B Eyepieces O Height adjustment condenser C Diopter adjustment I Kohler iris diaphragm D Nosepiece J Collector lens E Objectives K iCare sensor F Stage with X Y mechanical stage L Light intensity adjustment knob G X Y stage controls M Coaxial coarse adjustment H Condenser with iris diaphragm I Kohler iris diaphragm J Collector lens A B A D E F G J I H K L M C N O ...

Page 5: ...ransport handle at the backside of the microscope and with the right hand the bottom of the microscope Transport handle Caution Holding the microscope with the stage the stage focusing knob will damage the microscope Insert the power cord in the backside of the microscope and use the cable storage CSS Cable Storage System to store the cable after use CSS Cable Storage System Caution If the bacteri...

Page 6: ...lied with the head assembled However if your order contains the fluorescence it should be mounted first The dovetail on the bottom side of these parts fits into the slot on the top side of the other parts Placing the eye pieces 1 Remove the cover of eyepiece tube 2 Insert the eyepiece into the eyepiece tube The eyeshades Each eyepiece has its rubber eyeshade This prevents damage to the lens and pr...

Page 7: ...ustment please find the screw for locking the eyepiece on the tube ring here A Please note that location can be slightly rotated from model to model For models with diopter adjustment take out the eyepiece and look into the tube to find the right position of screw B A B ...

Page 8: ...slide with the cover glass facing up 3 Rotating the X and Y axis knob will move the specimen to the center for alignment with the center of the objective Focusing and slide protection 1 Select the objective 4x to the optical path 2 Rotate the position screw to top observe the right eyepiece with right eye Rotate the coarse focusing knob until the image appears 3 Rotate the fine focusing knob for d...

Page 9: ...distance from the eyepiece to the user s pupil To obtain the correct eye point move the eyes towards the eyepieces until a sharp image is reached at a full field of view Adjusting the diopter Using a binocular or trinocular tube is less tiring for the eyes than the use of a monocular tube In order to obtain the right interpupillary setting one should go through the below steps Turn the diopter adj...

Page 10: ...bed to the field of view it shows condenser has been centered correctly The field Köhler diaphragm By limiting the diameter of the beam entering the condenser the field diaphragm can prevent other light and strengthen the image contrast When the image is just on the edge of the field of view the objective can show the best performance and obtain the clearest image The diaphragm is factory pre cent...

Page 11: ...of the objective moves in the oil and the bubbles will disappear 10 After using the S100x objective turn the table with the fine adjustment knobs downwards until the front lens doesn t touch the oil any longer 11 Always clean the front lens of the S100x objective with a piece of lens paper that is moistened with a drop of isopropanol We recommend using Euromex lens paper isopropanol 12 Clean the s...

Page 12: ...ntrast Slider 1 Keep the phase contrast slider face up text up insert it from left to right into the condenser slider socket as the direction of the arrow pointed 2 Each slider has 3 positions 2 phase contrast positions and in the center of the slide the bright field position for normal use without phase contrast Each phase contrast objective used has to be matched with the phase contrast ring on ...

Page 13: ... bright field For phase contrast the condenser position should match the objective used Meaning that when the condenser is in position 40 the objective used should also be 40x Rotate 10 infinity plan phase contrast objective into the field of view then set the condenser to match the objective marker 10 20 Take the eyepiece out of the tube and insert centering telescope in its place Observed from c...

Page 14: ...roscope cleaning kit PB 5275 It is not necessary and not recommended to clean the lens surfaces at the inner side of the objectives Sometimes dust can be removed with high pressured air There will never be dust in the objectives if the objectives are not removed from the revolving nosepiece Caution Cleaning cloths containing plastic fibers can damage the coating of the lenses Maintenance of the st...

Page 15: ...e the fuse from the cover and examine the fuse If the thin piece of metal going from one end of the fuse to the other has a gap then the fuse is bad 6 If the fuse is bad install a replacement fuse in the cover 7 Gently push the fuse cover with the new fuse back into the sub woofer until it is flush with the unit Turn the cover clockwise about 3 4 to secure the cover back into the unit Note Fuse ma...

Page 16: ...t adapter into the 23 2mm tube of the phote port For focussing slowly unscrew the tube A you will be able to match parfocality of the camera with the view through the eyepieces by moving the camera up and down inside the 23 2mm tube Take an easy to view specimen and focus the image through the microscope s eyepieces with dioptre adjustment set on 0 Afterwards perform the height adjustment procedur...

Page 17: ...S 7210 Plan PLi 20x 0 25 IOS objective infinity corrected IS 7240 Plan PLi S40x 0 65 IOS objective infinity corrected IS 7260 Plan PLi S60x 0 85 IOS objective infinity corrected IS 7200 Plan PLi S100x 1 25 IOS objective infinity corrected IS 7710 Plan Phase PLPH 10x 0 25 objective IS 7720 Plan Phase PLPH 20x 0 40 objective IS 7740 Plan Phase PLPH S40x 0 65 objective IS 7700 Plan Phase PLPH S100x 1...

Page 18: ...Page 18 IS EN V4 ...