background image

User's Guide 

 

 

 

 

 

Model 412355

A

 Current / Voltage Calibrator 

 
 
 
 
 
 
 

Summary of Contents for 412355A Oyster Series

Page 1: ...User s Guide Model 412355A Current Voltage Calibrator ...

Page 2: ...C adaptor jack 9 Battery compartment 10 Neck strap holders 11 Function switch Battery Replacement When the LOW BAT message appears on the LCD replace the 9V battery as soon as possible 1 Open the calibrator s lid as far as possible 2 Open the battery compartment using a coin at the arrow indicator 3 Replace the battery and close the cover You as the end user are legally bound Battery ordinance to ...

Page 3: ...f activated FUNCTION SLIDE SWITCH Slide the Function switch on the side of the meter to the desired function Voltage or Current MODE Press the MODE button to select either SOURCE output or MEASURE input UNIT BUTTON Press the UNIT button to select 1 Voltage mode mV or V in source 2 Current mode mA or in source or measure OUTPUT ADJUST BUTTONS Press the buttons to increase or decrease the output vol...

Page 4: ...n the current mode or too low in the voltage mode Storing values into memory 1 Select and display a memory location M1 to M5 2 Press the buttons to adjust the display to the desired source value 3 Press and hold the MEM button for 2 seconds The displayed value will be stored into the displayed memory location Default memory values Five common source values are permanently programmed in memory as t...

Page 5: ...ODE button to select SOURCE 4 Press the UNIT button to select V or mV if voltage is selected or to select mA or if current is selected 5 Connect the Calibration Cable to the meter 6 Connect the Calibration Cable to the device or circuit under test 7 Use the buttons to select the desired output value Use the LCD display to verify the output level Alternatively use the stored calibration values in m...

Page 6: ...ting Altitude 2000meters 7000ft maximum Dimensions 96 x 118 x 45mm 3 8 x 4 7 x 1 8 folded Weight 340g 12 oz Accessories Supplied 9V battery AC adaptor and calibration cable with spade lugs Range Specifications Mode Function Range Resolution Accuracy of reading Measure Voltage 0 to 2000mV 1mV 0 075 1 digit or 3 digits whichever is greater 2 00 to 20 00V 0 01V Current 0 00 to 50 00mA 0 01mA 25 0 to ...

Reviews: