Operation   of   the   FEI   Tecnai   F30   Polara   

(22/03/2016)   

   

Initial   Checks   
   

1.

 

Ensure   the   following   software   applications   are   all   running   and   open   in   

this   order:   
-­‐

 

Tecnai   User   Interface   

-­‐

 

Filter   control   

-­‐

 

Digital   Micrograph   

-­‐

 

TEM   Imaging   and   Analysis   (TIA)   

If   you   need   to   restart   the   software   close   all   programs   and   restart   in   the   
above   order.   Sometimes   logging   out   and   back   in   again   can   sort   out   bugs.   
The   username   is   supervisor   and   the   password   is   f30d316   

2.

 

Check   the   vacuum.   The   Gun   should   be   at   Log   1   and   the   Column   at   Log   6.   
The   Camera   should   be   below   Log   35.   

3.

 

Make   sure   the   High   Tension   is   ON   (High   Tension   button   Yellow)   and   the   

FEG   is   ON   (Operate   button   is   Yellow).   

4.

 

Check   the   LN2   levels   in   the   inner   and   outer   dewars.   The   inner   dewar      (He   

dewar)   should   be   between   13-­‐14   (left   display)   and   the   outer   dewar   
should   be   above   70%.   The   inner   dewar   needs   topping   up   every   2-­‐3   days.   
The   outer   dewar   will   empty   at   a   rate   of   ~5%/hour   and   should   not   go   

below   20%.   

5.

 

Use   the   metal   funnel   to   top   up   the   Insertion   Rod   dewar.   This   dewar   can   

last   up   to   10   hours.   If   the   Insertion   Rod   warms   up   the   vacuum   
deteriorates   and   the   HT   may   switch   off.   When   collecting   with   EPU   a   LN2   
top-­‐up   at   11pm   will   ensure   it   stays   cold   until   9am   the   next   day.   

6.

 

Check   the   temperatures   of   the   different   channels   on   the   display:   
1.   Helium   cryo   shield   
2.   Nitrogen   cryo   shield   

3.   Helium   cold   finger   
4.   Nitrogen   cold   finger   

5.   Guide   tube   
6.   Insertion   rod   
7.   Selection   rod   

7.

 

If   the   specimen   Selection   Rod   is   still   attached   to   the   microscope,   move   it   
to   the   transfer   station   

-­‐   Make   sure   column   valves   are   closed.   
-­‐   Vent   airlock   
-­‐   Under   Setup   -­‐>   Helium   -­‐>   Vent   airlock   

-­‐   Wait   for   message   to   close   valve   B   
-­‐   Close   valve   B   (Pull   out   a   bit,   then   push   all   the   way   in   to   lock)   
-­‐   Wait   for   message   to   close   valve   A   

-­‐   Close   valve   A   (Pull   out   a   bit,   then   push   all   the   way   in   to   lock)   
-­‐   Detach   the   temperature   lead   via   the   USB   connection   

-­‐   With   one   hand   on   the   plastic   manipulator   at   the   base,   release   
clamps,   rotate   and   lift   off   hinges,   move   over   to   the   transfer   station   
-­‐   Place   on   hinges,   rotate   flush,   fasten   clamps   

Summary of Contents for Tecnai F30 Polara

Page 1: ...ur and should not go below 20 5 Use the metal funnel to top up the Insertion Rod dewar This dewar can last up to 10 hours If the Insertion Rod warms up the vacuum deteriorates and the HT may switch off When collecting with EPU a LN2 top up at 11pm will ensure it stays cold until 9am the next day 6 Check the temperatures of the different channels on the display 1 Helium cryo shield 2 Nitrogen cryo ...

Page 2: ...ition should be accessible 18 Remove old grids and C clips from the cartridges using 5 tweezers 19 Record the bottom and top cartridge identifier digits on the notepad 20 Begin cooling the transfer chamber initially without the lid and then through the side funnel to avoid condensation inside Continue cooling the chamber and keeping the LN2 level just above the cartridge loading area 21 After a co...

Page 3: ... to dock Secure the top clamp first followed by the bottom clamp 41 Clock on Pump Airlock 42 V43 will open followed by V41 Once P10 gets to Log 36 the message to open VA will appear 43 Pull VA to open and wait for the pins to pop in 44 A prompt will shortly follow asking to open VB 45 Pull VB to open and wait for the pins to pop in The Selection Rod is now connected to the Microscope vacuum Turn t...

Page 4: ...rds to the final stop and ensure the silver button pops up Once locked double check by gently trying to move the insertion rod to ensure it is locked in 61 On the holder push in the white knob until the red stripe disappears and then screw clockwise one complete turn to capture the threads on the cartridge Continue turning until finger tight DO NOT OVERTIGHTEN The Cartridge button should turn Yell...

Page 5: ... ...

Page 6: ...Spirits one session is up to half a day 4 hours for negative stain and up to a full day 8 hours for cryo EM A half day is for example 9am 1pm 1pm 5pm or 5pm 9pm A whole day would be 9am 5pm You can of course book less time if you don t need the full 4 or 8 hour slot The actual booking times are also flexible e g 10 12pm Please note that evenings and weekends count the same on the Spirits If on the...