FEI Tecnai G

F20 Operating Procedures 

                       (6/01/2017) 

   

   

1. Startup

 

   

   

   

1.

 

Sign-up 

in the microscope log-book 

2.

 

Startup the software if not already running: 

Start the software in this 

sequence: 

   

1.  

Tecnai User Interface

 

2.  

Digital Micrograph

 

 

3.  

TIA

 

(TEM Imaging and Analysis)  

   

3.

 

Check the vacuum

: Look at the 

Vacuum 

panel under the 

Setup 

menu. The 

GUN, COLUMN and CAMERA values should look like the ones below and 
highlighted in green. Gun will always be 1 Log. 

 
4.

 

Fill the cold-trap dewar with LN

2

Cover the screen window first. Fill the dewar 

until it is full. If you are the first user of the day, refill it again after 30 minutes. 
Refill the LN

at least every 3 hrs to keep it full 

 

   

 

   

 

 
 

5.

 

Inserting the specimen holder  

 

Room temperature sample 

There are two things to check prior to inserting the specimen holder: 
(a) Under the 

Search tab

, in the 

Stage

2

 menu, click on the flap-out menu and click 

Control

 then 

Holder

. This will reset X, Y, Z, and 

α

 on the stage and avoid damaging the 

Summary of Contents for Tecnai G2 F20

Page 1: ...the ones below and highlighted in green Gun will always be 1 Log 4 Fill the cold trap dewar with LN2 Cover the screen window first Fill the dewar until it is full If you are the first user of the day...

Page 2: ...ding Remove the objective aperture by sliding the lever to the right Under Setup tab click on Turbo to switch on the turbo pump The turbo pump will start speeding up orange button color and in 3 minut...

Page 3: ...ol panel 2 Centering the condenser aperture Condenser a Select a C2 aperture 100um for TEM Single particle b At a magnification of 10 000 go to crossover and center the beam using the left trackball c...

Page 4: ...e phosphor screen using Multifunction knobs X and Y b Beam Tilt Pivot Points X and Y Click on Beam Tilt Pivot Points X and go to crossover Try to minimize the movement of the beam using Multifunction...

Page 5: ...sert the objective aperture in slide left and using the aperture position knobs on the aperture center it around the center of the un diffracted beam 6 Objective lens stigmation To stigmate the object...

Page 6: ...k Search or Preview to view the image and click Focus and move the red box to a contrasty area to adjust focus using the objective focus knob i Click Acquire to obtain an image In the TIA analysis mod...

Page 7: ...the 50K_recommended lds file or set up your low dose as below At 50 000x the pixel size is 2 08 Search Focus Exposure 2 sec integration time Magnification Detector 1 700x Orius 150 000x Orius 50 000x...

Page 8: ...Appendix Microscope Information Line resolution nm 0 144 Cs objective mm 2 0 Cc objective mm 2 0 Focal length mm 2 7 Detector FEI Falcon II Physical pixel size m 14...