background image

1

55XXA/COIL 50, 55XXA/COIL 10

50, 1, 2, and 10-Turn Current Coil

Instructions

Introduction

Use the 55XXA/COIL 50, 55XXA/COIL 10 (the Product or Coil) to calibrate clamp-type current 

meters (clamp meters) that operate by two different principles - as current transformers (ac 

only), and by the Hall Effect (both ac and dc). For example, it is impractical to calibrate 

1000 A-rated current clamp meters with a 1000 A source. However, you can effectively 

multiply the current of the current source calibrator by a factor of 50 to support the 

calibration and verification of these clamp meters with the 50 turns of the 55XXA/COIL 50.

Contact Fluke

Fluke Corporation operates worldwide. For local contact information, go to our website: 

www.flukecal.com

.

To register your product, or to view, print, or download the latest manual or manual 

supplement, go to our website.

+1-425-446-5500  

fluke-info@fluke.com

Fluke Corporation

P.O. Box 9090

Everett WA 98206-9090

U.S.A.

Fluke Europe B.V.

P.O. Box 1186

5602 BD Eindhoven

The Netherlands

PN 5227455
April 2022

 Rev.1, 12/22

 

(English)

©2022 Fluke Corporation. All rights reserved.
Specifications are subject to change without notice.
All product names are trademarks of their respective companies.

Summary of Contents for 55XXA/COIL 50

Page 1: ...r of 50 to support the calibration and verification of these clamp meters with the 50 turns of the 55XXA COIL 50 Contact Fluke Fluke Corporation operates worldwide For local contact information go to our website www flukecal com To register your product or to view print or download the latest manual or manual supplement go to our website 1 425 446 5500 fluke info fluke com Fluke Corporation P O Bo...

Page 2: ...e Product if it is altered or damaged Symbols Table 1 lists the symbols used on the Product or in this document Table 1 Symbols Symbol Description Consult user documentation W WARNING RISK OF DANGER X WARNING HAZARDOUS VOLTAGE Risk of electric shock P Conforms to European directives This product complies with the WEEE Directive marking requirements The affixed label indicates that you must not dis...

Page 3: ... output Specification of calibrator current 0 13 Specification due to the Coil 0 65 The RSS of these two specifications determines the total specification of the source coil combination For higher frequencies or higher current levels the compliance voltage may exceed 1 V If the compliance voltage exceeds 1 V then compliance adders must be added to the Calibrator specification 55XXA COIL 50 55XXA C...

Page 4: ...l see Figure 1 and Figure 2 55XXA COIL 10 see Figure 1 For 10 turns center the clamp meter on the Coil base For 1 turn and 2 turns align the clamp meter with the markings on the Coil base 55XXA COIL 50 see Figure 2 Center the clamp meter on the base Calibration accuracy to specifications is guaranteed only with proper clamp alignment Align the clamp meter as much as possible on the base during cal...

Page 5: ...50 1 2 and 10 Turn Current Coil Use the Coil 5 Figure 2 55XXA COIL 50 CO MM ON 10 TU RN 30V ...

Page 6: ...luke authorized service center to obtain return authorization information then send the product to that Service Center with a description of the problem THIS WARRANTY IS YOUR ONLY REMEDY NO OTHER WARRANTIES SUCH AS FITNESS FOR A PARTICULAR PURPOSE ARE EXPRESSED OR IMPLIED FLUKE IS NOT LIABLE FOR ANY SPECIAL INDIRECT INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES ARISING FROM ANY CAUSE OR THEORY Sin...

Reviews: